JINHE XDV-300E Thermal deformation/Vicat softening point temperature tester, digital dial meter, touch screen, computer operation
Digital dial meter, touch screen, computer operation, used to measure the thermal softening performance of plastic raw materials or finished products. By changing the indenter, the instrument can measure the thermal deformation temperature and Vicat temperature, and supports the thermal deformation and Vicat softening point of various samples specified by the standard. Thermostatic pretreatment, highlighting Linear dispersion display, report editing
-
JINGKELIAN LR-2 asphalt softening point tester
$ 59.83 -
INESA WQD-1A Drip point softening point tester
$ 1396.09 -
JINHE XDV-300A Thermal deformation/Vicat softening point temperature tester, digital dial meter, touch screen
$ 1973.14 -
INESA WQD-400 automatic video softening point meter, room temperature~ 400 ℃
$ 4422.18 -
YOUKE YK3656 Hot deformation Vicat softening point testing machine
$ 3794.61
XDV-300E Heat Deformation·Vicat Softening Point Temperature Tester adopts the new international and national standard design, which is used to measure the heat softening performance of plastic raw materials or finished products. By changing the indenter, the instrument can measure the heat distortion temperature and Vicat temperature, using transformer oil (within 200 ℃) or methyl silicone oil (within 300 ℃) as the heat conduction medium, mechanical stirrer for oil circulation in the box, large capacity Oil tank design, stable temperature control and uniform temperature field.
main feature
1. Support the thermal deformation test of various samples specified in the standard.
2. Support various Vicat softening point tests specified in the standard.
3. The dual-control mode of the upper and lower computers can be controlled by the microcomputer, and the lower computer touch screen is used to control when the computer is not connected.
4. Support constant temperature pretreatment and static pretreatment mode.
5. Graphic curve data display interface.
6. Perfect test report editing function.
7. Test items and environment configurations are saved and opened independently, which is very convenient to use.
Standard
GB/T1633-2019 "Determination of Vicat softening point temperature of thermoplastics"
GB/T1634-2004 "Plastics - Determination of load deformation temperature"
GB/T8802-2001 "GB T 8802-2001 Vicat softening temperature of thermoplastic pipes and fittings"
JinHe XDV-300E Microcontrol thermal deformation Vicat softening point temperature testerSpecifications
LIST | VALUE |
---|---|
heating rate | 50 ℃/H and 120 ℃/H |
Temperature measurement | PT100 Sensors, center point measurement |
Temperature resolution | 0.1℃ |
Maximum temperature error | ±0.5℃ |
Deformation measurement method | Digital dial or dial meter |
Deformation measurement range | 0-12.7mm (can be configured according to user requirements) |
Maximum error of deformation measurement | 0.005mm/0.0005mm |
Deformation measurement resolution | 0.01mm/0.001mm |
Structural form | desktop |
Number of sample racks | 3 (can be customized according to user requirements) |
Hot deformation test span | 64mm、100mm |
Heating medium | Methyl silicone oil (suitable for room temperature -300 ℃, viscosity requirements below 100CS) or transformer oil (used within 200 ℃) |
Fuel tank capacity | About 13 liters |
Heating power | 3KW |
Power | 220V ± 10% 15A 50Hz |
Chill down | Natural chilling down or circulating water cooling |
Net weight | 70 kg |
JinHe XDV-300E Microcontrol thermal deformation Vicat softening point temperature tester Packing list
complete-built unit X1, manual X1, certificate of conformity X1, warranty card X1
[Note] Because the manufacturer's packaging may be updated or upgraded, the detailed packaging list shall be subject to the latest standard configuration of the manufacturer.
FAQ
- ActiveISO
- ActiveGB/T 20221-2023
- ActiveGB/T 10002.2-2023
- ActiveGB/T 10002.1-2023
- ActiveGB/T 22374-2018
- ActiveGB/T 15332-1994