HANNA HI934 Karl Fischer Coulomb Method (KF) Potentiometric titrator
Moisture content ranging from 1 ppm to 5%, resolution 0.1 ppm, dynamics with 3 speed settings, statistical mean of results, standard deviation, powerful built-in algorithms for termination criteria based on fixed mV endpoints or absolute/relative drift.

-
HANNA HI933 Karl Fischer Volumetric (KF) Moisture Meter Potentiometric titrator
$ 9330.41 -
LEICI ZDJ-4B Potentiometric titrator
$ 3692.09 -
YueFeng ZD-2A Automatic Potentiometric titrator 0~ 14.00pH
$ 343.43 -
HANNA HI931 Automatic Potentiometric titration, poten-tiometry analyzer
$ 7989.41 -
HANNA HI934 Karl Fischer Coulomb Method (KF) Potentiometric titrator
$ 9330.41
Introduction
The HI934 is a Karl Fischer coulomb titrator with high accuracy, flexibility and repeatability. This titrator is designed to perform titrations for a variety of applications, allowing the user to obtain good results and high-speed analysis. The moisture content range analyzed by HI934 is 1ppm to 5%. This powerful titrator efficiently monitors KF reactions, detects endpoints, and performs all necessary calculations and plots.
Features
Small footprint, requiring minimal bench space;
Housing made of sturdy, chemical-resistant plastic;
Powerful built-in algorithms for termination criteria based on fixed mV endpoints or absolute/relative drift;
Sample analysis averages and statistics;
Minimize water vapor ingress by sealing the solvent system;
Balance interface for automatic weighing;
Supports 100 titration methods;
User-defined reports;
Warning and error messages are clearly displayed.
HANNA HI934 Potentiometric titratorSpecifications
LIST | VALUE |
---|---|
Measurement range | 1 ppm to 5% |
resolution | 0.1 ppm |
unit of result | %,ppm,mg/g,μg/g,mg,μg,mg/ml,μg/ml,ppt,mgBr/100g,gBr/100g,mgBr,gBr |
Sample category | Liquid or solid (externally dissolved or extracted) |
pre-titration condition | automatic |
Background drift correction | Automatic or user-selectable |
endpoint metric | Fixed mV continuous, relative drift stop or absolute drift stop |
Measurement | Dynamic with 3 speed settings |
result statistics | Mean, standard deviation |
Titration container type | Borosilicate Glass with Standard Cone Glass Joints |
operation volume | 100 - 200 ml |
Titration vessel diaphragm | silicone rubber |
Diaphragm cap thread | GL-18 |
reagent port | Standard taper 19 |
Electrode type/connection | HI76330 |
Glass connection | Standard taper 14/20 |
Polarization current | 1,2,5,? 10μA |
voltage range | 2 - 1100 mV |
voltage resolution | 0.1 mV |
Accuracy @25 ℃ | ±0.1% |
Generator electrode model | Diaphragm or no diaphragm |
Detection electrode type | automatic |
Electrical connection | 5-Pin connector with detachable cable |
Glass connection | Standard taper 29/12 |
Maximum current | 400 mA |
current control | Automatic or fixed (400mA) |
Type of mixing equipment | Magnetic, Electronic Regulation, Digital Stirrer |
Stirring speed | 200 - 2000 RPM |
Stirring resolution | 100 RPM |
External Stirrer | 4-Pin mini-DIN connection allows control of external mixing equipment |
Type of reagent processing system | A sealing system with integrated diaphragm air pump |
Type of drying agent | molecular sieve |
Bottle thread type | GL-45 |
Glass connection | Standard taper 19 (using adapter provided) |
Reagent/Waste Tube | PTFE |
show | 5.7 "Graphic Color Display with Backlight |
peripheral equipment | PC (USB Standard B); Flash Drive (USB Standard A); Analytical Balance (DB-9 socket); Printer (DB-25 socket); Keyboard (6-pin Mini-DIN) |
language | English |
Power Consumption | 100-240 VAC,50/60 Hz / 0.5 Amps |
Plate material | ABS/PC and stainless steel |
keyboard | Polyester brazing |
Operating conditions | 10 - 40 0C;80% RH |
Preserving the environment | -20 - 70 0C;95% RH |
Size/weight | 315 x 205 x 375 mm / 4.3kg |
HANNA HI934 Potentiometric titrator Packing list
machine X1, manual X1, certificate X1, warranty card X1
[Note] Because the manufacturer's packaging may be updated or upgraded, the detailed packaging list shall be subject to the latest standard configuration of the manufacturer.
FAQ
- ActiveGB/T 3653.1-2024
- ActiveGB/T 43788-2024
- ActiveSN/T 2261-2009
- ActiveHB 20056.4-2011
- ActiveYB/T 159.3-2015
- ActiveYB/T 505.4-2007
- Active 178-2014
- ActiveYS/T 710.1-2009
- Active 5110.2-2000
- Active 0995-2019
- Active 0980-2019
- ActiveSH/T 0688-2000
- ActiveHG/T 3838-2008
- ActiveYC/T 314-2009
- ActiveSH/T 1767-2008
- Active 0836-2010
- Active 5092.4-2001
- Active
- ActiveSN/T 4951-2017
- Active 6060-2022