Anmiao AT-140S (Neutral) -CN Ultrasonic Tachymeter Operation Instructions

1. Features

* Suitable for soft materials such as wood and rubber.

* Using dedicated microcomputer LSI and crystal time base to provide high-precision measurement.

* With high energy emission and broadband reception sensitivity, this meter can match probes of different frequencies. This makes it easy to measure rough surfaces, even cast iron. Widely used in various industries.

* Automatic shutdown to save power.

* It can communicate with PC computer through optional RS232C interface cable and software for statistics and printing.

2. Performance parameters

Display: LCD

Speed ​​range: 500~9000 m/s

Resolution: 10 m/s

Accuracy: 0. 5%±0. 5% n

Setting thickness: 0. 70~50. 00mm

Power supply: 4 AAA batteries

Operating conditions:

Temperature: 0~50℃

Humidity: <80%RH

Dimensions: 140×70×30 m

Weight: about 130 g (without battery)

Standard configuration:

Carrying case. . . . . . . . . . . 1 pc.

Instruction Manual . . . . . . . . . . 1 copy

ultrasonic sensor

An Miao AT-140S (Neutral)-CN Ultrasonic Sound Velocity Meter Operation Instructions with Figure 1

4. Measurement procedure

4.1 Take a sample of known thickness.

4.2 Press the "power button" to turn on the machine.

4.3 Press the "thickness key" to input the known thickness into the measurement table. After pressing the "thickness key", the thickness can be displayed on the display. Any known thickness can be changed from 0.80 to 50.00 mm by pressing the plus or minus key. The variable is 0.01 mm each time the plus or minus key is pressed; and, if pressed for more than 4 seconds, the variable is 0.1 mm; if pressed for more than 8 seconds, the variable is 1.00 mm; if pressed for more than 12 seconds, the variable is 10.0mm. Changed thicknesses are automatically saved in the table for later use.

4.4 Apply a little couplant on the material to be tested, and press the sensor well on the surface. If the coupling is good, the coupled symbol lights up. Shows the sound velocity reading of the measured material on the display. The reading will be held until the next measurement. The last value will be held until the power is turned off.

4.5 The instrument has two shutdown methods, manual shutdown by pressing the power button at any time, and automatic shutdown without pressing any button within 10 minutes.

4.6 How to turn off the buzzer sound? To turn off the beep, just press the plus and minus keys at the same time. Turn on the buzzer, and press the plus and minus keys at the same time again.

5. Calibration

5.1 Apply a little oil to the 5mm standard block.

5.2 Clamp the standard block with two probes, press the calibration key, the display will show the reading "2960 m/s" and the symbol "CAL".

6. Replace the battery

6.1 When the battery symbol appears on the display, it is time to replace the batteries.

6.2 Slide the battery cover to remove the battery from the instrument.

6.3 Pay attention to the polarity when installing the battery.