Relation table between coating 4 cup viscosity and Miscellaneous viscosity cup

Viscosity Cup Test Instructions

Before the test, clean the inner surface of the viscometer with gauze soaked in solvent, dry it with cold air, and observe that the hole of the leak nozzle should be clean. Place the viscometer on a horizontal position, place a 100ml measuring cup under the viscometer, heat the adhesive sample to 25+0.5°C in advance, block the leak hole of the viscometer with your left middle finger, and quickly pour the sample into the viscometer cup , Use a glass rod to scrape air bubbles and excess samples into the groove, so that the liquid level in the cup is flush with the edge of the cup. If the room temperature is higher than 25°C, control the sample at 25-0.5 in advance. Release the tightly pressed left middle finger to let the adhesive flow out, and start the stopwatch at the same time to record the time (s) when the flow of the adhesive flows out of the leak hole of the viscometer.

Tu 4 cup viscosity and other Viscosity Cup relationship table

Absolute viscosity/Pa.sTu-4 cup viscosity/sKinematic viscosity/cm².s-1Tu-1 cup Viscosity/sEngler viscosity condition4# Ford cup viscosity/s
0.050180.514.56.9420
0.065twenty two0.666.08.9326
0.085280.857.011.534
0.100301.017.513.640
0.125321.218.016.346
0.140381.429.519.1751
0.165421.6910.122.8257
0.18045/11.0/60
0.200502.0212.027.2765
0.225572.2914.030.9275
0.250652.5616.034.5685
0.275732.7818.037.5396
0.300802.9720.040.09108
0.320893.2122.043.34117
0.3401233.4231.046.17123
0.3701283.7132.050.09127
0.4001334.0033.054.00131
0.4351384.3634.059.86137
0.3000.3000.3000.3000.3000.300
0.470144
   147
4.74/63.99144
0.5001545.04/68.04147
0.5501665.45/73.58154
0.627/6.69/90.32166
0.884/8.95/120.83/