Thermo320P-01 Portable pH meter Manual Operating instructions

introduce

This method provides a method for measuring the pH value of conventional water. Using the 9107BNMD three-in-one pH electrode, it can be widely used in the field measurement of the pH value of conventional water samples.

Thermo320P-01 Portable PH Meter Manual Operating Instructions with Figure 1

Preparation of Calibration Standards

1. Make sure all buffers are at the same temperature.

2. Tear off the pH7.00 buffer solution bag and put it away for later use.

3. Tear off the pH4.01 buffer bag and put it away for later use.

4. Tear off the pH10.01 buffer bag and put it away for later use.

sample preparation

1. Measure 100mL sample with a graduated cylinder and pour it into a 150mL beaker.

2. If you need to measure other samples, please repeat the above steps.

electrode storage

In order to ensure fast response of the electrode and smooth flow of the liquid junction, do not allow the sensitive glass bulb and liquid junction to dry out.

Short-term storage of electrodes (less than one week)

Soak the electrode in 910001pH electrode storage solution. If there is no such storage solution, please add 1g KCl to 200ml pH7 buffer solution as a temporary replacement.

Long-term storage of electrodes (more than one week)

Put a few drops of 910001 electrode storage solution into the electrode protective cap, and then gently cover the electrode to protect the liquid junction and sensitive glass bulb. When using the electrode again, precondition it as if it were a new electrode.

Equipment preparation Electrode preparation

1. Gently remove the black glass bulb protective sleeve from the electrode and put it away for later use.

2. Clean the white deposited salt on the outside of the electrode with deionized water.

3. Gently shake the electrode, (just like shaking a thermometer), to remove the air bubbles in the electrode.

4. Soak the electrode in pH electrode storage solution 910001 for one hour. If there is no electrode storage solution, 1g KCl can be added to 200mL pH7 buffer as a temporary electrode storage solution.

Thermo320P-01 Portable PH Meter Manual Operating Instructions with Figure 2

Thermo320P-01 Portable PH Meter Manual Operating Instructions with Figure 3