1. Product introduction:
LD-5 electric centrifuge is suitable for hospital laboratories, clinics, and biochemical laboratories. The qualitative analysis is used for plasma, serum, urea, and vaccine manufacturing.
2. Technical parameters:
1. Power supply: 220V50HZ
2. Maximum speed: 4000r/min
3. Maximum relative centrifugal force: 2395g
4. Capacity: 10×50ml,
5. Noise: <75dB(A)
6. Timing range: 0~60 minutes/minute (digital speed measurement)
7. Dimensions: 35×31×28(cm)
3. Safety instructions:
It is strictly forbidden to run without load without installing the rotor; check the rotor for corrosion damage before operation; do not touch the running rotor when the rotor is running; strictly prohibit the operation of the rotor when it is loaded unbalanced.
4. Instructions for use:
The workbench of the centrifuge should be flat and firm, and the working room should be neat, clean, dry and well ventilated, and the ambient temperature should be 5-32°C. Turn on the centrifuge and wipe the inner cavity and rotor clean. Put the test tubes with consistent weighing , close the centrifuge cover, adjust the speed control knob, and increase to the required speed. Before each stop, the speed control knob needs to be placed at the minimum position. "Timer" is set to "zero", turn off the power switch, and cut off the power. Wipe the inside Cavity and rotor, close the centrifuge cover.
Five, maintenance:
Do not place any objects on the centrifuge cover. After each use, be sure to clean and wipe the inner cavity and rotor, and cover the centrifuge cover to prevent dust and dirt from entering. If the centrifuge has not been used for a long time, it should be cleaned before it is used again. Open the cover for a period of time to dry the inner cavity. The carbon brushes of the motor should be replaced in time after 3000 hours of use to avoid wear on the commutator. After a long-term use of the centrifuge, natural wear and tear is a normal phenomenon. After one year of use, if there is a breakdown Or in case of difficulties, the company will continue to provide good services and assist in handling them.