1. Product introduction:
LD-4 low-speed centrifuge is suitable for hospital laboratories, clinics, and biochemical laboratories for qualitative analysis of plasma, serum, urea, and vaccine production.
2. Technical parameters:
1. Power supply: 220V50HZ
2. Host speed: 4000r/min
3. Maximum relative centrifugal force: 2800g
4. Capacity: 4×100ml or 10×15ml5. Noise: <75Db(A)
6. Weight: 49kg
3. Safety instructions:
It is strictly forbidden to run without load without installing the rotor; check the rotor for corrosion damage before operation; do not touch the running rotor when the rotor is running; strictly prohibit the operation of the rotor when it is loaded unbalanced.
4. Instructions for use:
The workbench of the centrifuge should be flat and firm, the working room should be neat, clean, dry and well ventilated, and the ambient temperature should be 5-32°C. Turn on the centrifuge and wipe the inner cavity and rotor clean. Put the test tubes with the same weight, close the centrifuge cover, adjust the speed control knob, and increase the speed to the required speed. Before each shutdown, the speed control knob needs to be placed at the minimum position. Cut off the power. Wipe the inner cavity and rotor, and close the centrifuge cover.
Five, maintenance:
Do not place any objects on the centrifuge cover. After each use, be sure to clean and wipe the inner cavity and rotor, and cover the centrifuge cover to prevent dust and dirt from entering. If the centrifuge has not been used for a long time, it should be removed before it is used again. Open the cover of the centrifuge for a period of time to dry the inner cavity. The carbon brushes of the motor should be replaced in time after 3000 hours of use to avoid wear and tear on the commutator. In case of failure or difficulty, the company will continue to provide good service and assist in handling it.