KQ-100 ultrasonic cleaner use steps

1. The minimum amount of cleaning liquid added to the cleaning tank shall not be less than 30mm. The maximum liquid level , and the space for the items you need to add should also be taken into account.

2. Put the items to be cleaned into the cleaning grid, and then put the cleaning grid into the cleaning solution. Never put the items directly into the cleaning tank, so as not to affect the cleaning effect and damage the instrument.

3. Insert the plug of the ultrasonic cleaner into the 220V power socket (grounding device is required to use the power supply).

4. Press the ON power switch, the green switch power indicator light will be on, indicating that the power supply is normal and can work .

5. Press the ON heating switch, the temperature indicator light will be on, the heater has been heated to 70°C, the temperature indicator light will go out, the heater will stop working, when the temperature drops below 50°C, it will automatically heat up, and the temperature indicator light will be on .

6. When the heating temperature reaches 70°C, the working time of the cleaning timer can be turned on. The position of the timer can be adjusted arbitrarily within 0-20 minutes, and can also be adjusted to the normal position: the general cleaning time is 10-20 minutes, and it can be used for particularly difficult-to-clean objects Appropriately extend the cleaning time.

7. After cleaning, take out the rack or hanger from the cleaning tank, and spray it with warm water or rinse it in another solvent-free warm water cleaning tank.

8. After rinsing, dry, store and assemble.