The instruction manual of the trace moisture analyzer comprehensively introduces the use, assembly and maintenance process and steps of the instrument in detail, but for the sake of safer and long-term use of the customer, the installation site of the trace moisture analyzer is introduced as follows:
1. The instrument shall not be installed in a place exceeding the specified power fluctuation.
2. The instrument shall not be installed near frequently operated electrical equipment.
3. The instrument shall not be installed where the room temperature is lower than 5°C or higher than 40°C.
4. The trace moisture analyzer shall not be installed in a room with corrosive gas, the corrosive gas may corrode the circuit part of the instrument and shorten the life of the instrument.
5. The instrument must not be installed in a place exposed to direct sunlight.
6. The instrument must not be installed in places with high humidity.
7. In order to make the trace moisture analyzer work more reliably, it is recommended to use an AC voltage stabilizer or UPS power supply.