Moisture meter electrolysis solution how to replace

The trace moisture analyzer is a kind of instrument that is often kept in the laboratory. In its manual, the operation, assembly and maintenance of the instrument have been comprehensively introduced, but these are basic operations. In normal use, only pay attention to maintenance , this problem can be avoided. In order to ensure the accuracy of the determination of the trace moisture analyzer, it is usually necessary to pay attention to the replacement of the electrolyte.

Before replacing the electrolyte, prepare a clean filter paper, then put the two drying tubes in the electrolytic cell on the paper, then take out the electrolytic electrode, pour out the electrolyte inside it, and then put it in the electrolytic cell Pour out the electrolyte solution slowly. It is worth noting that: in the process of pouring the electrolyte solution, do not pour out the stirring bar.  

Under normal circumstances, there is no need to clean the electrolytic cell, but if it is really polluted, it can be cleaned, but it must be dried after cleaning. During the cleaning process, pay special attention to the white plate at the bottom of the electrolysis electrode, because it is a honeycomb ceramic plate, once it contains water, it is difficult to be released, so try not to use water for cleaning.

There is really no way, when you have to use clean water to clean, after cleaning, put it in an oven with a temperature of about 50°C and dry it. The time can be controlled at about 6 hours, and then let the white plate at the bottom of the electrolytic electrode of the trace moisture analyzer Cool naturally in the oven.