A constant magneticstirrer operation precautions

A constant magnetic stirrer operation precautions with Figure 1

1 Environmental requirements for this product: the altitude does not exceed 4000 meters, and the temperature is between 5°C and 40°C for indoor use; it needs to be placed on a flat and stable table, at least 25cm away from the wall or other objects, and at least 80cm away from the top of other objects;


2 Please use this product in a laboratory environment. The personnel who use and operate this product need to undergo professional study or training, understand laboratory safety knowledge, and have certain basic common sense, professional knowledge and related experiments in physics, chemistry, biology, medicine, etc. room skills.


3 To use this product, it needs to be connected to an appropriate power supply, and the power supply used needs to be effectively grounded;


4 Stirring or heating needs to be liquid or soluble powder that does not affect the safety of the experiment: it cannot be flammable, explosive, volatile, easy to solidify, or violently exothermic, or hazardous to refrigeration. reactions, etc.; if you need to use it, please be responsible for your own safety!


5 Wear safety glasses and appropriate protective equipment when operating this product;


6 Keep the environment clean and dry, away from strong heat sources, strong magnetic fields, flammable and explosive materials or other similar magnetic products;


7. This product cannot be immersed in liquid and cannot be moved during use. When moving or cleaning, the power cord needs to be unplugged;


8 The rated maximum stirring volume is 5 liters (H 2 O), the maximum load capacity of the Heating Plate does not exceed 12Kg, and the maximum stirring volume can reach It should be too full to avoid a dangerous situation due to overflow when stirring;


9 The Heating Plate (the outer surface of aluminum alloy is enamel or stainless steel) and the shell (the outer surface of aluminum alloy is sprayed with plastic paint) cannot be touched by sharp objects; any part of the body should avoid contact with any corrosive substances, please remove it immediately if it is accidentally touched; Cleaning: Use cotton cloth and water-soluble detergent, and then wipe dry; when the Heating Plate temperature is high during and after work, keep power cords, sensor leads and other objects away from them to prevent scalding and burning, and do not touch them to prevent burns ;


10 The container holding the solution should not be magnetically conductive or airtight, preferably a container with a thin and flat bottom of high-temperature-resistant tempered glass or non-magnetic stainless steel; the items of the operator need to be antimagnetic and need to be kept away from this product (Heating Plate There is a strong magnetic field inside);


11 There is a temperature sensor inside the Heating Plate. When the inside of the plate exceeds 350°C (or 370°C), it will stop heating and other work to play a protective role. Some models will sound a buzzer to prompt, press "+" or " -” key can eliminate the alarm sound. It can be reused . This machine cannot be used alone as an electric stove or other heating appliances;


12 Fuse specifications: 250V5A, need to match when replacing, the position of the fuse box is under ;


13 This machine only controls the temperature of the Heating Plate, but the temperature of the heated liquid is uncontrolled and uncertain; if you want to accurately control the temperature of the stirred liquid, please choose IT-09A or IT-08A models (constant temperature Magnetic stirrer);


14 The machine has been tested before leaving the factory. The surface of the Heating Plate may be slightly discolored, but it does not affect the function. Please use it with confidence.