BINDER oven common faults and corresponding solutions


Fault description

possible reason

required measures


temperature


The set point temperature is not

Arrived after the specified time.

incorrect door

closure.

Close the chamber door completely.





The door gasket was defective.

replace the doormat,


Controller is not adjusted.

Calibrate and adjust controllers.


Wrong voltage.

Check if the power supply is correct

Voltage (Chapter. 4.2).






BF, FD, Fed: Fans Not

Turning or turning too slowly.

BF, Feeding: Fan speed set too low

BF, FEED: Set fan speed to 100%






The fan is defective.

Contact Binder Service.





Controller is defective.

Contact Binder Service.


room heating for a long time,

The Pt 100 sensor is defective.


Set points are not held.

Semiconductor Relay Defects



Controller is not adjusted.

Calibrate and adjust controllers.


The room was not hot. hot

Display the icon.

Defective heating element.

Contact Binder Service.



The semiconductor relay is defective.


The room was not hot. No

heating icon in the display. fraud

Controller display works.

timer ran off

Reprogram the timer or turn it off.


Semiconductor relay defect.

Contact Binder Service.


Controller is defective.



non-functional room,

The standby icon is displayed.

The pod is in standby mode.

Press the standby button until


The display lights up.




BD, BF: Alarm message "!TProt "

will display

Safety equipment class 3.1 has

respond.

Check temperature settings


True set point and safe elimination



Sub 3.1 class (Chapter. 7.3).





ED, FD, Fed: No in the chamber

Function. The warning message "!TLim" will be displayed

Safety equipment class 2 has

Closed the room.

Let calm down the chamber. examine

Temperature setting -

Point and Safeguard Classes

2 (Chapter. 7.3). Select if appropriate

Appropriate limit value.









Safety device class 2 defec-

up.

Contact Binder Service.





The room has no function.

There is no power.

Check connection to power.


Overheating protection

Device class 1 is closed

conference hall.

disconnect the chamber from

Power on for at least 10 seconds

Let it cool down. if the device

Responsive again, contact binder

Serve






Controller is defective.

Contact Binder Service.


Deviation from indicated

heating time.

The room is fully loaded.

room for less or consider

Extend heating time.





controller


The message "1999" in the controller

show

Sen-

and controller.

Contact Binder Service.