NGP8 Series Portable Intelligent Gas Detector Product Manual

Instructions

1 Key definition: There are three keys on this machine, ⊙ key, ↑ key, ↓ key (order from left to right)

             The ⊙ key is the power switch and the confirmation function, the ↑ key is the shift key, and the ↓ key is used for page turning and digital addition and subtraction.

2   boot method

Press and hold the ⊙ key for 5 seconds, and it will automatically turn on after 5 seconds. After turning on, the screens displayed in sequence are: gas name, maximum range, unit and power. After 2 seconds, the preheating countdown will start to display for 120 seconds. If you don’t want to Wait, press the ⊙ key to directly enter the normal measurement interface, the gas type is displayed in the lower left corner, and the vibration motor, buzzer, and indicator light all alarm once to indicate normal startup. If the battery is insufficient or the sensor cannot be detected, the alarm will continue. , at this time, if you press the ⊙ key to silence the sound, the sound, vibration motor, and indicator light will stop working, but the word "Alarm" is still flashing on the screen. If the battery is insufficient, the battery icon will flash. If the sensor cannot be detected , and the word "FAULT" is displayed , press the ⊙ key again at this time, the sound, vibration motor and indicator light will continue to work. If the instrument is a pump suction type, the pump icon will be , and it is always rotating, if there is no pump, the icon will not be displayed. The picture on the right is the screen displayed after normal startup (take CO as an example)

NGP8 Series Portable Intelligent Gas Detector Product Manual with Figure 1NGP8 series portable intelligent Gas Detector product manual with picture 2NGP8 series portable intelligent Gas Detector product manual with picture 3NGP8 series portable intelligent Gas Detector product manual with picture 4

Menu Settings

Password setting: press ⊙ key, ↑ key, ↓ key in sequence, and it will be valid only if it is completed within 2 seconds. Menu definition: Z, S, FS, AL, AH, CER, DEFA, OUT

1. Zero calibration Z

If the zero point drift of the sensor is too large, zero point calibration is required, as shown in Figure 1

 NGP8 series portable intelligent Gas Detector product manual with picture 5

The zero point setting menu defaults to "N", as shown in Figure 2, and it needs to be confirmed whether to operate.

NGP8 Series Portable Intelligent Gas Detector Product Manual with Figure 6

If it is confirmed to calibrate the zero point, in the zero point setting menu, press the ↓ key to modify it to "Y", as shown in Figure 3  

NGP8 Series Portable Intelligent Gas Detector Product Manual with Figure 7

Press the ⊙ key to confirm the zero point calibration, if the calibration is successful, the word "YES" will appear in the lower left corner, as shown in Figure 4

NGP8 Series Portable Intelligent Gas Detector Product Manual with Figure 8

If the calibration is unsuccessful, the word "NO" will appear in the lower left corner, as shown in Figure 5 

NGP8 Series Portable Intelligent Gas Detector Product Manual with Figure 9

6.2. Target point calibration S 

If the sensor has been used for a long time, sensitivity correction is required , use the ↓ key to select the " S " option, and press the ⊙ key to modify it, as shown in Figure 6    

NGP8 Series Portable Intelligent Gas Detector Product Manual with Figure 10

The target point concentration setting menu defaults to "N", as shown in Figure 7, and it needs to be confirmed whether to operate.  

NGP8 Series Portable Intelligent Gas Detector Product Manual with Figure 11

If it is confirmed that the target point needs to be calibrated, in the target point setting menu, press the ↓ key to modify it to "Y", as shown in Figure 8

NGP8 Series Portable Intelligent Gas Detector Product Manual with Figure 12

Press the ⊙ key to confirm the calibration of the target point. If the calibration is successful, the word "YES" will appear in the lower left corner, and the value will change to the value set in the menu "CER", as shown in Figure 9 

NGP8 Series Portable Intelligent Gas Detector Product Manual with Figure 13

If the calibration is unsuccessful, the word "NO" will appear in the lower left corner, as shown in Figure 10 

NGP8 Series Portable Intelligent Gas Detector Product Manual with Figure 14

Full scale setting FS

As shown in Figure 11, use the ↓ key to select the "FS" option, press the ⊙ key to modify, and the ↑ key to shift, and then press the ⊙ key to save after modification.

NGP8 Series Portable Intelligent Gas Detector Product Manual with Figure 15

Level 1 alarm point setting AL

As shown in Figure 12, use the ↓ key to select the "AL" option, press the ⊙ key to modify, and the ↑ key to shift, and then press the ⊙ key to save.

NGP8 Series Portable Intelligent Gas Detector Product Manual with Figure 16

Secondary alarm point setting AH

As shown in Figure 13, use the ↓ key to select the "AH" option, press the ⊙ key to modify, and the ↑ key to shift, and then press the ⊙ key to save after modification.

NGP8系列便携式智能气体检测仪产品说明书配图17

Target point concentration setting CER

As shown in Figure 14, use the ↓ key to select the "CER" option, press the ⊙ key to modify, the ↑ key to shift, and then press the ⊙ key to save.

NGP8系列便携式智能气体检测仪产品说明书配图18

Restore factory default DEFA

As shown in Figure 15, use the ↓ key to select the "DEFA" option, press the ⊙ key to modify, and you need to confirm whether to operate.

NGP8系列便携式智能气体检测仪产品说明书配图19

Exit menu settings OUT

As shown in Figure 17, use the ↓ key to select the "OUT" option, press the ⊙ key to set and exit

NGP8系列便携式智能气体检测仪产品说明书配图20

equipment maintenance

In normal use of the Detector, the effective service life of the sensor is 24-36 months. During the effective service life, the sensor should be regularly calibrated and checked every 6 months or 1 year to ensure accurate and effective gas monitoring. Outdated and faulty sensors need to be replaced.

sensor replacement

After the sensor fails, please return the instrument to the factory for replacement.

Sensor Calibration

   See 6.2 description for details. Do not operate when the standard gas is not ready. In case of misoperation, you can refer to the description in 6.7 to restore the factory default settings. Refer to the description in 6.2 before calibration, set the concentration value to be calibrated in advance, and then pass the standard gas, the flow rate of the standard gas is controlled at 300 ml/min

Precautions

l It is strictly forbidden to charge or disassemble the instrument on site

l It is strictly forbidden to replace the sensor with power on

l The calibration check of the Detector should be carried out regularly

l Sensors that have exceeded the effective service life and have faults should be replaced in time

l Avoid impacting the sensor with a gas higher than the measurement range, which may damage the sensor, and man-made damage is not covered by the warranty.  


NBCHAO E-shop
Guess you like it
Relevant knowledge