Huazhi HZY/PTY/DTY series electronic analysis balance installation method

HZY, PTY, DTY series electronic balances can be widely used in medical scientific research institutions, schools, enterprises, institutions, highway construction and defense departments, etc., and are satisfactory instruments for high-speed and accurate weighing.

unpack

1) Check the balance for visible damage immediately after opening the box

2) Before the balance is successfully installed, please keep all packing materials for return shipping. Before packing the balance , please remove all connecting cables to avoid unnecessary damage

Packing List

Huazhi HZY/PTY/DTY series electronic analytical balance installation method with diagram 1

Installation Notes

When choosing a location for the balance, please note the following:

- Do not place the balance near heating or direct sunlight, where the air flows directly (open windows or doors)

——Do not place the balance in an environment where the temperature is too high or too low, prone to collision, vibration and humidity

——Installed on a level, stable, and vibration-free workbench, and shall not be placed in places with rot and turbidity or strong magnetic field interference places that affect the normal operation of the balance

Adapting the Balance to Temperature Changes

When moving a balance at a lower temperature to a place with a higher temperature or at a higher temperature to a place with a lower temperature, please leave the balance at room temperature for about 2 hours, and then turn it on and preheat it for use . The balance is aligned with the new ambient temperature equilibrium.

Installation steps

Huazhi HZY/PTY/DTY series electronic analytical balance installation method with diagram 2

Huazhi HZY/PTY/DTY series electronic analytical balance installation method with picture 3

Huazhi HZY/PTY/DTY series electronic analytical balance installation method with diagram 4

Huazhi HZY/PTY/DTY series electronic analytical balance installation method with picture 5Use dry batteries/rechargeable batteries (only for optional balances)

○Dry batteries or rechargeable batteries are not included in the scope of delivery of the balance

△ Only common 9V dry batteries or rechargeable batteries on the market can be used

Only use an external charger to charge the rechargeable battery

Place the balance on its side

Open the battery cover

Put in 9V dry battery or rechargeable battery

Confirm that the positive and negative poles of the battery are installed correctly

Used dry batteries or rechargeable batteries are special garbage (not household garbage). According to the Waste Disposal Act, rechargeable batteries are collected as special waste and disposed of separately.

Close the battery compartment: press the battery compartment cover down and tighten the screws





Huazhi HZY/PTY/DTY series electronic analytical balance installation method with diagram 6Level the balance

The balance must be rebalanced every time the balance is changed. To level the balance, just use the two rear anchor bolts.

Turn the last two anchor bolts counterclockwise into place

Turn the rear anchor bolt as shown until the bubble of the level is exactly in the center of the ring.

Turn the last two anchor bolts clockwise to the supporting surface

>Typically, the leveling procedure needs to be repeated several times.