Coverage rate Tester instrument parameters

    Coverage Tester instrument parameters

function code

parameter name

Factory default

Remark

F — 00

Enter password

00000

"05282" modify parameter secret word "01266" parameter initialization secret word

F — 01

 sampling mode

    1

"0" continuous sampling mode

"1" One-shot mode

F — 02

Operating mode

0

"0" debug mode

"1" working mode

F — 03

Battery voltage alarm value setting

1.2000 V

Setting range 0 - 1.5 V

F — 04

2# standard plate correction coefficient

1.0000

Setting range 0 - 1.9999

F — 05

High reflectivity correction factor

1.0000

Setting range 0 - 1.9999

F — 06

Medium (upper) reflectivity correction factor

1.0000

Setting range 0 - 1.9999

F — 07

Medium (lower) reflectivity correction factor

1.0000

Setting range 0 - 1.9999

F — 08

Low reflectivity correction factor

1.0000

Setting range 0 - 1.9999

F — 09

Calibration plate reflectance standard value

82.00

Setting range 0 - 99.99

F — 10

High reflectivity standard value

92.50

Setting range 0 - 99.99

F — 11

High reflectivity actual value

00.00

Setting range 0 - 99.99

F — 12

Middle (upper) reflectance standard value

82.58

Setting range 0 - 99.99

F — 13

Middle (lower) reflectance standard value

76.32

Setting range 0 - 99.99

F — 14

Actual value of middle (lower) reflectivity

00.00

Setting range 0 - 99.99

F — 15

Low reflectivity standard value

58.55

Setting range 0 - 99.99

function code

parameter name

Factory default

Remark

F — 16

Low reflectivity actual value

00.00

Setting range 0 - 99.99

F — 18

Zero offset

00.00

Setting range 0 - 99.99

F — 17

Calibration plate standard value error range

15.00

Setting range 0 - 99.99

F — 18

Instrument serial number

00000

Setting range 0 - 99999

illustrate:

1. When the correction coefficient of 2# standard plate is set to 1.0000, it means that the 2# standard plate will not be corrected.

2. 1# standard board reflectivity=92.50 standard high reflectivity

    2# standard plate reflectance = 82.58 reference value

3# standard plate reflectance = 76.32 standard medium reflectance

4# standard plate reflectance=58.55 standard low reflectance

High reflectivity correction factor applies to 100.0 => correction factor > 87.54

    Applicable range of mid (upper) reflectance correction factor 87.54 => correction factor > 79.45

Applicable range of middle (lower) reflectance correction factor 79.45 => correction factor > 67.43

Low reflectivity correction factor applicable range 67.43 => correction factor > 0.0 

3. When the actual value of high reflectivity, the actual value of middle (lower) reflectivity and the actual value of low reflectivity are set to zero, the nonlinearity will not be corrected.