Why can't the salt spray Test Chamber be heated?

During the use of the instrument, some failures are unavoidable. For example, the non-heating of the salt spray Test Chamber is a common problem. We have made the following solutions to the temperature problem of the salt spray Test Chamber.

1. The salt spray Test Chamber is not heated

Reasons and solutions:

1. The heating solid state relay has no output, and the solid state relay needs to be replaced

2. The instrument has no output, and the instrument needs to be replaced

3. The heating tube is damaged, replace the heating tube

4. The lack of water leads to heating shielding, check whether the water level is normal, and whether the floating ball has malfunctioned

5. The over-temperature protection malfunctions, adjust the protection temperature value, if it still cannot be solved, please replace the over-temperature protector

2. The heating temperature of the salt spray Test Chamber is abnormal

Reasons and solutions:

1. The temperature setting of the temperature controller in the laboratory is too low, set the temperature controller to the required temperature

2. The setting of the safety protection switch in the laboratory is too low, set the safety protection switch to the required temperature

3. The temperature of the saturation barrel temperature controller is too low, set the temperature controller to the required temperature

4. The setting of the safety protection switch of the saturation barrel is too low, set the safety protection switch to the required temperature