Instructions
| 1, | Plug in the power plug and insert the 8MM air pipe into the compressed air inlet. |
| 2, | Turn on the power switch, and the low water level lights, low salt water lights, and low water level lights in the pressure tank above the control panel are all on. The lights indicate that the Test Chamber, the salt Water Tank, and the pressure tank are short of water. Good water as well as salt water. |
| 3. | The low water level light in the laboratory is on: it means that there is a lack of water in the laboratory, and water needs to be added (tap water is sufficient), until the low water level light in the laboratory goes out. |
| 4. | The low water level light of the pressure tank is on: it means that there is a lack of water in the pressure tank. The pressure tank is hidden in the distribution box and we cannot see it. Add water directly from the manual water filling place of the pressure tank. Before adding water, turn off a red ball valve at the back Open it and add it until the low water level light of the pressure tank goes out. After adding water, remember to close the red ball valve. (Tap water is enough) |
| 5. | The low salt water light is on: it means that there is no salt water in the salt Water Tank, we need to mix the salt water well and inject it from the reagent inlet. until the low salt water light goes out. (Saline configuration method is as follows) |
| 6. | Add water to the water seal tank until it can submerge the surrounding area of the transparent acrylic cover. |
| 7. | After all the water is added, the operation switch can be turned on below, and the temperature of the test room and the pressure tank will be displayed. The temperature of the test room is set at 35°C, and the temperature of the pressure barrel is set at 47°C. After the temperature is set, you can put the V-shaped shelf and the black O-shaped support rod, put the workpiece to be tested on these two layers, and close the cover. |
| 8, | After putting the product in place, you can turn on the spray switch. Once the spray switch is turned on, the pressure gauge will display the pressure. Adjust the spray pressure to 1Kg/cm² by adjusting the front pressure regulating valve. (Note: the spray pressure must be tens of millions Do not overpressure, do not exceed 1Kg/cm², overpressure is easy to burst the tube. The pressure set at the compressed air inlet is 2Kg/cm².) |
| 9, | First set the time to be tested on the timer, turn on the timing switch after counting the time, and automatically end the test when the time is reached. |
Salt water preparation method: 500g of sodium chloride in a bottle, mixed with 9.5L of purified water, and stirred evenly with a stick.
Maintenance
1. If the test period exceeds 1 month, please replace the water in the heating tank.
2. If the saline solution used in the test has not been used for more than 1 month, please do not use it again, so as not to affect the quality of the test.
3. If there is a long time interval before the next test, please clean the interior of the laboratory after the test and drain the water in the heating tank.
※Drain the water in the heating tank---------------Open the drain valve.
Isolate the water discharge in the sink---------------Pull up the middle silicone plug.
Drain the water in the preheating tank---------------Pull up the inner silicone plug.