Learn how to choose a Lepto thermostatic circulator in one article

internal application

First determine whether it is an internal application or an external application. Internal application refers to the test that the sample is placed in the bath for constant temperature, heating or cooling; external application refers to the application of cooling, heating or constant temperature of the external device, the external device may be Closed or open.

For internal applications, it is necessary to clarify the requirements of the sample, such as temperature range, sample type, sample weight and immersion depth, whether gradient temperature changes are required, whether it is required to reduce the temperature within a specified time or how much the temperature is raised, etc., according to the requirements of the sample. Determine the constant temperature The requirements of the circulator are mainly the following:

temperature range

temperature stability

Program control function, the test of some samples requires a temperature gradient change requirement during the test process, which requires the equipment to have a program control function. For program control applications, it is also necessary to consider whether the heating power and cooling power meet the requirements of sample heating and cooling. Require

Slope control function: For the test of how much the temperature needs to be lowered or how much the temperature needs to be raised within the specified time, the equipment needs to have the slope control function. For slope control applications, it is also necessary to consider the heating power and whether the cooling power meets the requirements for sample heating and cooling

Bath opening size: it is the key to ensure whether the user's sample can be put down

Bath depth: ensure that the sample volume in the test container used by the user can be fully immersed

Magnetic stirring: During sample testing and reaction, is it necessary for the equipment to provide a stirring function to achieve stirring of the sample?

Sink cover: According to the user's observation of the sample, confirm whether the cover is transparent, whether it is required to be arched, whether it needs to be changed in size during application, etc.

Vertical/horizontal: Horizontal machine is suitable for placing on the table, vertical is suitable for placing on the ground

Accessories: brackets, test tube racks, sensors, bath fluids, bath covers, adapters, etc. required during sample testing

external application

For external applications, the temperature control system needs to be connected to the application end through pipelines, typically such as jacketed reactors, etc. For such external applications, the following factors should be considered:

the nature of the material

temperature range

temperature stability

Distance between temperature control system and external application

Cycle Pressure Requirements

Circulation Flow Requirements

Heating Power: Determines the rate at which the external application heats up

Cooling power: Determines the rate at which the external application cools down

Whether external temperature control is required: accurate material temperature control is very necessary for users

Connection interface/pipeline specifications: ensure smooth installation

remote control

body wash

Function selection reference

typemodelCooling capacity
(W)
Heating
power
(W)
Temperature range
[°C]  
Temperature control accuracy
(°C)
Maximum lift
/rated
   (m)

Maximum flow /rated
L/min
Bath
volume
[L]  

Low Temperature Coolant
Circulator _




Minicooler200/ -5…35±25131
icooler-2003450/-20…RT±0.32.0/1.018/123
icooler-2005700/-20…RT±0.33.0/2.022/155
icooler-4005800/-40…RT±0.33.0/2.022/155
icooler-20091200/-20…RT±0.510/828/159
icooler-40091500/-40…RT±0.510/828/159
icooler-20172400/-20…RT±0.510/828/1517
icooler-30172400/-30…RT±0.510/828/1517
icooler-8005400/-80…RT±0.518/12Pressure 2.0/1.0
   Suction 1.0/0.5
5
High-low temperature
constant temperature
circulator
icooler-2003+4001500-20…100±0.12.0/1.0123
icooler-2005+7001500-20…100±0.13.0/2.022/155
icooler-4005+8001500-40…100±0.13.0/2.022/155
icooler-2009+12001500-20…100±0.110/828/159
icooler-4009+15002000-40…100±0.110/828/159
icooler-2017+24002000-20…100±0.110/828/1517
icooler-2003P4501500-20…100±0.02Pressure 0.2
   Suction 0.1
123
icooler-2006P7001500-20…100±0.02Pressure 0.3
   Suction 0.2
twenty two6
icooler-2009P12001500-20…100±0.021.0twenty two9
icooler-4006P8001500-20…100±0.02Pressure 0.3
   Suction 0.2
twenty two6
icooler-4009P15002000-20…100±0.021.0twenty two9
High-precision
chiller
(air-cooled type)
UC06600/5-35±0.18/517/67
UC08800/5-35±0.18/517/67
UC101000/5-35±0.142/155.3/4.07
UC121200/5-35±0.142/155.3/4.013
UC161600/5-35±0.142/155.3/4.013
UC222200/5-35±0.1200/408.0/7.513
UC303000/5-35±0.1160/4020/17.013
UC363600/5-35±0.127/2158.3/33.325
UC525200/5-35±0.127/2158.3/33.325
Chiller
(air-cooled type)
IC0101000/5-35±18/517/67
IC0121200/5-35±142/155.3/4.013
IC0161600/5-35±110/twenty two/7
IC0222200/5-35±127/2158.3/33.330
IC0272700/5-35±127/2158.3/33.317
IC0363600/5-35±127/2158.3/33.325
IC0525200/5-35±127/2158.3/33.325
High-low temperature
constant temperature bath
(vertical economical type)
CC-1008E3002000-10…100±0.05~0.12.0/1.0128
CC-2008E4002000'-20...100±0.05~0.12.0/1.0128
CC-2508E4002000-25…100±0.05~0.1Pressure 2.0/1.0
   Suction 1.0/0.5
128
CC-0513E2000 -5...100  ±0.05~0.11213
CC-1013E2000-10...100  ±0.05~0.11213
CC-2013E2000 -20...100  ±0.05~0.11213
CC-3013E2000-30…100±0.05~0.1Pressure 2.0/1.0
   Suction 1.0/0.5
1213
CC-0517E2000 -5...100  ±0.05~0.1twenty two17
CC-1017E2000-10...100  ±0.05~0.1twenty two17
CC-2017E2000 -20...100  ±0.05~0.1twenty two17
CC-0520E9002000-5…100±0.05~0.13.0/2.0twenty two20
CC-2020E9002000-20…100±0.05~0.13.0/2.0twenty two20
CC-3020E9002000-30…100±0.05~0.13.0/2.0twenty two20
CC-0527E9002000-5…100±0.05~0.13.0/2.0twenty two27
CC-2027E9002000-20…100±0.05~0.13.0/2.0twenty two27
High-low temperature
constant temperature bath
(vertical standard type)
CC-10083002000-10…200±0.01~0.052.0/1.0128
CC-20084002000-20…200±0.01~0.052.0/1.0128
CC-2013
2000 -20...220  ±0.01~0.05twenty two13
CC-3013






CC-2017
2000 -20...220  ±0.01~0.05twenty two17
CC-3017






CC-1020
200010...220  ±0.01~0.051220
CC-2020
2000 -20...220  ±0.01~0.05twenty two20
CC-2508 
2000 -25...200±0.01~0.05

CC-2508P2000 -25...250±0.01~0.05

CC-3020
2000 -30...220  ±0.01~0.05twenty two20
CC-2027
2000 -20...220   ±0.01~0.05twenty two27
High-low temperature
constant temperature bath
(horizontal economical type)
CC-0506EW2000-5…100±0.05~0.1126
CC-0508EW2002000-5…100±0.05~0.12.0/1.0128
CC-1008EW2002000-10…100±0.05~0.12.0/1.0128
CC-2008EW3002000-20…100±0.05~0.12.0/1.0128
High-low temperature constant temperature
(horizontal tank standard type)
CC-1008W2002000-10…150±0.01~0.052.0/1.0128
CC-2008W3002000-20…150±0.01~0.052.0/1.0128
Calibration bathCC-1008wPro2000 -10...100±0.01

8
CC-0513Pro2000 -5...100±0.01

13
CC-0517 Pro2000 -5...100±0.01

17
CC-1015Pro2000 -10...250±0.01

15

Special constant temperature bath for capillary viscometer
VTC18/2000(-20)20...100±0.05~0.11818
VTC26/2000(-20)20...100±0.05~0.11826
FTC10200/ -5…RT±2
/1
High temperature
constant temperature bath
BC206E/2000RT+10…100±0.05~0.12.0/1.0126
BC210E


±0.05~0.1
10
BC213E/2000 RT+10...100  ±0.05~0.12.0/1.01213
BC217E/2000RT+10…100±0.05~0.13.0/2.0twenty two17
BC220E/2000RT+10…100±0.05~0.13.0/2.0twenty two20
BC227E/2000RT+10...100  ±0.05~0.13.0/2.0twenty two27
BC320E/3000RT+10...100±0.05~0.1
20
BC327E/3000RT+10...100  ±0.05~0.1
27
BC206 /2000 RT+10...220±0.01~0.053.0/2.0126
BC208/2000RT+10…220±0.01~0.053.0/2.0twenty two8
BC210
2000RT+10…220±0.01~0.05

BC213
2000RT+10...220  ±0.01~0.052.0/1.01213
BC217
2000RT+10...220 ±0.01~0.053.0/2.0twenty two17
BC220
2000 RT+10...220 ±0.01~0.053.0/2.0twenty two20
BC227
2000RT+10...220 ±0.01~0.053.0/2.0twenty two27
BC320 
3000RT+10...220 ±0.01~0.05

BC327
3000RT+10...220 ±0.01~0.05

Open
thermostat
OBC206EA2000(-20)RT+10...100  ±0.05~0.1 126
OBC213EA2000(-20)RT+10...100 ±0.05~0.1  1212
OBC220EA/2000(-20)RT+10...100±0.05~0.1//20
OBC206A
2000(-20)RT+10...100±0.01~0.05
6
OBC213A
2000(-20)RT+10...100±0.01~0.05
13
OBC220A
2000(-20)RT+10...100±0.01~0.05
20
OBC206EB2000(-20)RT+10...100 ±0.05~0.1  126
OBC210EB2000(-20)RT+10...100 ±0.05~0.1  

OBC213EB2000(-20)RT+10...100  ±0.05~0.1 1213
OBC217EB2000(-20)RT+10...100 ±0.05~0.1  1217
OBC220EB/2000(-20)RT+10...100±0.05~0.1//20
OBC227EB2000  (-20)RT+10...100  ±0.05~0.1 1227
OBC327EB3000  (-20)RT+10...100  ±0.05~0.1 
27
OBC344EB3000  (-20)RT+10...100  ±0.05~0.1 
44
OBC358EB3000  (-20)RT+10...100  ±0.05~0.1 
58
OBC213B
2000(-20)RT+10...220 ±0.01~0.05
13
OBC220B
2000(-20)RT+10...220 ±0.01~0.05
20
OBC327B
3000(-20)RT+10...220 ±0.01~0.05
27
OBC344B
3000(-20)RT+10...220 ±0.01~0.05
44
OBC358B
3000(-20)RT+10...220 ±0.01~0.05
58


NBCHAO E-shop