Infrared frontal thermometer DT-8807S how to operate and precautions?

temperature measurement

The meter measuring probe is aimed at the measuring site (forehead/behind the earlobe/underarm/ear hole), keeping a distance of 1~4cm.

After pressing the measurement button for about one second, the beep will sound "beep" to indicate that the measurement is completed, and the LED screen will display the temperature reading. .

When the temperature measurement function of the instrument is set to the human body mode, "LO" will be displayed when the detected body temperature is lower than 35°C; " HI" will be displayed when the body temperature is higher than 42°C.

How to operate the infrared forehead thermometer DT-8807S and precautions?  Picture 1

Precautions for body temperature measurement

The correct operation method is the key to measurement accuracy, because infrared temperature measurement has higher requirements on the surrounding environment temperature, please pay attention to the following tips when using it:

  • When measuring body temperature, please point the thermometer at the center of the forehead (above the eyebrows and keep it vertical). The measurement site should not be covered by hair, sweat, hats, etc., and please confirm that the measurement distance is 1~4cm.

  • The person under test comes from a place with a large difference in ambient temperature (such as a low temperature outdoor environment and a high temperature indoor environment), and should be allowed to wait for 5 to 15 minutes .

  • Cold compresses on the forehead of patients with fever, sweating and other cooling measures will lead to low measurement results, and measurement in this case .

  • When the thermometer is taken out from a place with a large difference in ambient temperature and used, it should be placed in the measurement environment to stabilize for 20 to 30 minutes before use.

  • Do not measure body temperature in places with strong airflow, such as the air outlet of the air conditioner or under the fan.

  • Do not blow air on the temperature probe.

troubleshooting

wrong reasonreasonApproach
How to operate the infrared forehead thermometer DT-8807S and precautions?  Picture 2Ambient temperature changes too fastPlease place the product under the tested environment (15 ~ 35*C) for at least 30 minutes
How to operate the infrared forehead thermometer DT-8807S and precautions?  Picture 3The ambient temperature is outside the range of 15~35°CPlease place the product in the environment under test (15 ~ 35°C) for at least 30 minutes
How to operate the infrared forehead thermometer DT-8807S and precautions?  Picture 4Human body model: measure temperature above 42°CPlease use it within the measurable temperature range. If there is another error message, please call the customer service hotline
How to operate the infrared forehead thermometer DT-8807S and precautions?  Picture 5Human body model: measure temperature below 35°CPlease use it within the measurable temperature range. If there is another error message, please call the customer service hotline
Press the button and nothing happensThe battery is dead; the battery is reversedPlease replace the battery and install the battery correctly

    FAQ

Problem Descriptionpossible reasonsprocessing method
Measurements are lowUse after sweating, taking a bath, or applying cold compresses will lower the skin temperaturePlease wipe off the moisture and wait for 30 minutes to measure
The Detector is dirtyPlease use a soft cloth or cotton swab dipped in 70% alcohol to wipe the Detector lightly
Immediately measure cold weather from outdoors to indoorsPlease wait for the forehead temperature to stabilize before measuring
Not measuring correctly to the center of the foreheadPlease remove hair, hat, etc., and align the forehead measurement correctly
Measurements are highThe product is stored at low temperature, and it will be measured immediately at room temperature, which will make the measurement result very highPlease place the product in the use environment for 30 minutes before measuring
The product has been exposed to direct sunlightPlease place the product in the use environment for 30 minutes before measuring
after the subject criesAllow its temperature to be in a constant state before measuring
Measurements varyNot measured at the same locationPlease measure the position of the forehead
The measurement is different from the oral temperatureThere may be differences between the temperature value converted by the infrared forehead thermometer and the actual oral temperatureIt is recommended that family members take more regular measurements to know their normal body temperature

Maintenance and Cleaning

  • Do not get close to hot objects.

  • Use batteries within specifications, and do not mix rechargeable batteries with other batteries.

  • In order to prevent product failure, do not disassemble the product arbitrarily.

  • If the product is dropped or impacted, there will be errors in the test.

  • After the test, use neutral detergent and clean water to clean, then wipe the water with a soft cloth and put it in the battery box for storage.

  • Please take out the battery for storage when not in use for a long time.