The high-low Temperature Test Chamber is a high-precision instrument. During use, the staff need to operate strictly in accordance with the regulations. At the same time, regular maintenance is also required. If it is improperly operated or neglected to maintain, some failures will occur, which will affect the progress of the work.
The high-low Temperature Test Chamber is a testing equipment composed of multiple systems. Fault handling requires a comprehensive understanding and grasp of the equipment in order to quickly troubleshoot and handle faults. Below we introduce some practical solutions to common faults in high-low Temperature Test Chambers.
1. If the cooling speed of the high-low Temperature Test Chamber is slow or cannot be cooled at all, check the refrigeration system to check whether the compressor in the refrigeration system is damaged, whether the circuit is abnormal, whether the refrigerant is used upward or leaking, and whether the pipeline is blocked, and repair it in a targeted manner and replacement equipment. If the high-low Temperature Test Chamber has frost or ice on the condenser, heating can be used to volatilize the frost attached to it, and then it can be used normally.
2. For the problem of heating and humidification of the high-low Temperature Test Chamber machine, check whether the heating wire, humidification tube or corresponding solid state relay is damaged, and deal with it accordingly.
3. In the rainy season or in places with a lot of rain, due to the presence of water vapor on the door body and the sealing strip of the high-low Temperature Test Chamber, the air will be discharged. The door and sealing strip of the high-low Temperature Test Chamber can be cleaned, or some lubricating substances, such as talcum powder, can be applied to make the parts smooth and not easy to generate water vapor.
4. The high-low temperature test box presses the power button, but the high-low temperature test box does not respond and is not turned on. In this case, check whether the voltage of the high-low temperature test box is stable, whether the line is connected well, and whether there is any abnormality, and ask the electrician to solve it. The compressor also has certain requirements on the ambient air temperature. If it is winter and the temperature is below 5 degrees, it is necessary to turn on the air conditioner or use other heating equipment to ensure the temperature conditions required for the normal operation of the compressor.
5. Increased noise is also a common problem of high-low Temperature Test Chambers. Check the motor, fan, compressor and other components, and tighten the screws.
If you still can't find the reason after the above troubleshooting for the high-low Temperature Test Chamber, you should contact the manufacturer or technical personnel for professional treatment in time.