Note: The installation location should consider the heat dissipation rate of the constant temperature and humidity Test Chamber and the characteristics of easy inspection and maintenance
There should be a distance of more than 60 cm between the constant temperature and humidity Test Chamber and the wall and any other machines.
It should be placed on a flat, vibration-free ground (check with a level).
The ambient temperature should be maintained between 10°C and 30°C, 70±10%RH, in order for the machine to operate stably; if the ambient temperature changes too drastically (for example, within a few minutes by 5°C or more), the temperature The control and temperature drop cannot be controlled very stably.
The constant temperature and humidity Test Chamber should be kept away from heat sources and flammable/explosive substances.
The constant temperature and humidity Test Chamber should not be exposed to direct sunlight, and maintain indoor air circulation.
Please avoid setting it in dirty and dusty places.
Electric lines and drainage pipes should be shortened as much as possible.