How to solve the leakage of large constant temperature and humidity Test Chamber?

1. Check the waterway to see if the water inlet valve is too large, resulting in too much water. If so, just turn off the water valve;

2. Check whether the laboratory replenishment time setting is too frequent, resulting in excessive replenishment. If yes, just increase the replenishment interval and reduce the replenishment frequency;

3. Check the floor to see if the inclination angle of the thermostatic chamber is flat, because the inclination angle is also a factor that causes the water in the wet bulb to be unbalanced, so that water drops down. If there is such a situation, we only need to move the equipment to flat ground for use;

4. Check whether the water level tank on the right side of the laboratory is too high, resulting in too much water. At this time, the position of the water level tank needs to be adjusted. When adjusting the water level tank, it is recommended to communicate with professional technical engineers or purchase from the manufacturer, and then take corresponding measures.