What are the reasons for the slow cooling of the low Temperature Test Chamber?

When the low Temperature Test Chamber is used improperly, it may cause some failures, such as rapid temperature changes, slow cooling of the low Temperature Test Chamber, etc. 

1. First, consider the ambient temperature conditions of the low-Temperature Test Chamber. Both the contract and the quotation emphasize that the operating temperature of the equipment should work below 25 ℃, and the laboratory should be ventilated to maintain air circulation, but some customers may not care about the equipment operating at 35 ℃ Working above the temperature, close the laboratory, which will inevitably lead to slow cooling. In addition, long-term operation of equipment at high temperatures will cause aging and damage to the refrigeration system and electrical components.

2. Secondly, the refrigerant in the low temperature laboratory leaked. The freezing medium is considered the blood of the refrigeration system. If any leak point of refrigerant leakage occurs in any part of the refrigeration system, the cooling capacity will be low and this will naturally affect the cooling of the equipment. But you have to find a professional test box manufacturer to check this problem.

3. Then consider whether the refrigeration system of the low temperature test room is blocked. If the refrigeration system is blocked for a long time, it will cause great damage to the equipment and cause the compressor to be damaged.