The cleaning effect of an ultrasonic cleaner is not necessarily proportional to (power × cleaning time). Sometimes with low power, it takes a long time to remove the dirt. If the power reaches a certain value, sometimes the dirt will be removed quickly. If the power is too large, the cavitation intensity will be greatly increased, and the cleaning effect will be improved, but at this time, the more precise parts will also have corrosion spots, which is not worth the loss, and the cavitation phenomenon at the vibration plate at the bottom of the cleaning cylinder is serious, causing water point corrosion will also increase. When using organic solvents such as trichlorethylene, there is basically no problem, but when using water or water-soluble cleaning solutions, it is easy to be corroded by water spots. If the surface of the vibrating plate has been scarred, the cavitation corrosion at the bottom of the water is more serious under strong power, so the ultrasonic power should be selected according to the actual use.

Ultrasonic cleaners can be used to clean many things these days, but how do you clean them properly?
Clean endoscopic biopsy forceps and auxiliary instruments, as well as residual mucus after using endoscopic accessories, etc.
1. The power supply and power supply of the electric heater need to have a good grounding device.
2. It is strictly forbidden to start when there is no water or solvent in the tank, which will cause air vibration and cause the vibrating head to be scrapped or damaged.
3. For cleaning equipment with a heating system, it is strictly forbidden to turn on the heating switch when there is no liquid.
4. It is strictly forbidden to hit the bottom of the cleaning cylinder with heavy objects (iron pieces) to avoid damage to the energy converter chip.
5. The power supply of the ultrasonic generator should use a 220V/50Hz power supply alone, and be equipped with a voltage regulator above 2000W.
6. The bottom of the cleaning cylinder should be rinsed regularly, and there should be no excessive debris or dirt.
7. During the operation of the washing machine, do not put your fingers in the washing tank, otherwise you will feel tingling or discomfort.
8. Every time a new liquid is replaced, the parts can be cleaned after the ultrasonic wave is started.
9. Use clean water or aqueous solution as a cleaning agent, and it is forbidden to use alcohol, gasoline or any flammable gas as a cleaning agent to add to the washing machine. The company is not responsible for the resulting fire and explosion.
10. The method with better ultrasonic cleaning effect. The temperature of the ultrasonic cleaning tank is 3050°C. Choose the correct cleaning agent according to different cleaning objects. Cleaning agents are generally divided into water-based (alkaline) cleaning agents, organic solvent cleaning agents, and chemical reaction cleaning agents. Generally, water-based cleaners are mostly used, and different cleaning times are selected according to the degree of pollution of the object to be cleaned and the nature of the dirt.