FANGRUI CH1006 (N) Thermostatic Bath Temperature range RT + 10~100 ℃ Low level protection
The volume of the CH1006 (N) Thermostatic Bath of the FANGRUI instrument is 6L, the Display resolution is 0.1 degrees Celsius, the internal and external pump circulation mode is adopted, the CH1006 (N) Thermostatic Bath is a round opening, with the function of breaking the couple protection, and the liner is stainless steel.
FangRui-
SDPTOP CH1006 Super Thermostat + 5~ 95 ℃/6L Internal and External Circulation
$ 385.00 -
NELE DC-3010 Thermostatic Bath horizontal 10L/-30 ℃ +/- 0.05 ℃
$ 1269.00 -
Shangtian Jingyi DC-5010 Low Temperature Thermostatic Bath Digital thermostatic bath -50 ℃/10L
$ 4672.00 -
Qiwei HS-601A Super constant temperature Water Bath 20L/0.1 ℃
$ 318.00 -
BOZHEN DK-8D thermostatic sink, three holes
$ 371.00
Introduction
Fangrui Instrument CH1006 (N) thermostatic bath is a microcomputer temperature control constant temperature Water Bath (Oil Bath), using a new high-performance monolithic microcomputer control, self-tuning PID adjustment, using imported temperature sensor temperature measurement, high temperature control accuracy, small fluctuation, using imported stirring motor, low noise and long life. Using a large-screen backlit LCD screen, the parameters are clear and clear, beautiful and fashionable. The temperature control of the instrument is accurate, stable and reliable, and the operation is convenient and safe.
Features
1. Adopt high-performance single-chip microcomputer control, PID adjustment and self-tuning control technology, with high temperature control accuracy and short stability time;
2. It has the function of disconnection protection to prevent the instrument from getting out of control;
3. The upper limit over-temperature alarm function can be set by yourself, and if the range is exceeded, it will alarm immediately;
4. The instrument works stably and reliably, and the operation is very convenient;
5. Large-screen backlit LCD, the parameters are clear and clear, beautiful and beautiful;
6. The Circulating Chiller can lead the constant temperature liquid in the tank to establish a second constant temperature field;
7. The inner tank and table top are made of high-quality stainless steel, which is corrosion-resistant and easy to clean.
Application
CH1006(N) thermostatic bathThe constant temperature test can be carried out directly in the tank, or the external circulation function can be used to guide the controlled constant temperature liquid in the tank to other equipment (such as viscometer, density meter, tensiometer, polarimeter, rotary evaporator, Spectrophotometer , etc.) through a hose connected to other equipment.
CH1006(N) thermostatic bathIt is widely used in bioengineering, food, chemical industry, metallurgy, chemical analysis, petroleum and other fields, providing users with a high-precision, controlled, and uniform temperature constant field source, and is a satisfactory constant temperature device for research institutes, colleges and universities, factory laboratories, and quality inspection departments.
Product display diagram

FangRui CH1006(N) Thermostatic BathSpecifications
| LIST | VALUE |
|---|---|
| Temperature range | RT+10~100℃ |
| Temperature Fluctuation | ±0.05℃ |
| Temperature Resolution | 0.1℃ |
| Tank Volume | 6L |
| Tank opening Dimensions | 145*145*150mm |
| Temperature control mode | Microcomputer temperature control, self-tuning PID adjustment |
| Circular mode | Internal/external circulation |
| Heating power | 1000W |
| Dimensions | 340*190*370mm |
| Flow rate of pump | 10L/min |
| Net Weight of the whole machine | 7Kg |
| Low level protection | Yes. |
FangRui CH1006(N) Thermostatic Bath Packing list
host x1, cover x1, silicone tube x1, power cord x1, manual, certificate of conformity, warranty card x1.
[Note] Because the manufacturer's packaging may be updated or upgraded, the detailed packaging list shall be subject to the latest standard configuration of the manufacturer.
FAQ
- Instructions for assembly and use of Thermostat
- How to assemble and use Thermostat
- Viscosity Determination of Eccentric Falling Ball Determination
- 0.01 ℃ High Precision Low Temperature Thermostat
- Thermostat structure and working principle introduction
- ActiveGB/T 18856.4-2008
- ActiveFZ/T 50010.3-2011
- ActiveHG/T 3660-1999
- ActiveJB/T 9518-1999
- ActiveGB 9278-2008
- ActiveGB/T 5561-2012


CH1006(N)