FANGRUI instrument HDC1506L high-precision thermostatic circulating bath Temperature range -15~ 100 ℃
FANGRUI instrument HDC1506L high-precision thermostatic circulating bath adopts small vertical structure design, large LCD screen, HDC1506L high-precision thermostatic circulating bath has an effective volume of 6L and a temperature control accuracy of +/- 0.01 ℃.
FangRui-
NELE DC-60100 Thermostatic Bath 100L/-60 ℃ +/- 0.05 ℃
$ 12350.00 -
FANGRUI DC1506N (W) Horizontal low temperature constant temperature Water Bath Temperature range -15~ 100 ℃
$ 755.00 -
SDPTOP DCY-2015 Low Temperature Thermostat -20~95 ℃/15L
$ 1820.00 -
NELE NLSC-50 Super Constant Temperature Bath (Water/Oil), Circulating Bath (Water/Oil) 50L/100 ℃
$ 1019.00 -
CNSHP DC-0515 Digital Display Thermostatic Bath -5 ℃/15L
$ 1182.00
Introduction
Fangrui Instrument HDC1506L high-precision Constant Temperature Circulating Bath is a microcomputer temperature control constant temperature Water Bath (Oil Bath) using a new high-performance monolithic microcomputer control, self-tuning PID adjustment, using imported temperature sensor temperature measurement, high temperature control accuracy, small fluctuation; using good refrigeration compressor (some models are all imported brand compressors or compressor units), high refrigeration efficiency, fast cooling, low noise; using good fully sealed stirring motor, lubricating grease is not volatile, low noise and long life.
Features
1. Adopt high-performance single-chip microcomputer control, PID adjustment and self-tuning control technology, with high temperature control accuracy and short stability time;
2. It has the function of disconnection protection to prevent the instrument from getting out of control;
3. There is a liquid level protection device, which can effectively prevent various failures caused by the empty burning of the heating tube;
4. The upper limit over-temperature alarm function can be set by yourself, and if the range is exceeded, it will alarm immediately;
5. The instrument works stably and reliably, and the operation is very convenient;
6. The inner tank, table top and all parts immersed in liquid are high-quality stainless steel, which is corrosion-resistant and easy to clean;
7. Two sets of light-emitting digital tubes are used to display, and the set temperature and actual temperature are displayed separately.
Application
HDC1506L high-precision thermostatic circulating bathThermostatic tests can be carried out directly in the tank, or the controlled thermostatic liquid in the tank can be directed to other equipment by means of a hose connected to other equipment using the external circulation function.
HDC1506L high-precision thermostatic circulating bathIt is widely used in bioengineering, food, chemical industry, metallurgy, chemical analysis, petroleum and other fields, providing users with a high-precision, controlled, and uniform temperature constant field source, and is a satisfactory constant temperature device for research institutes, colleges and universities, factory laboratories, and quality inspection departments.
Product display diagram

FangRui HDC1506L Thermostatic BathSpecifications
| LIST | VALUE |
|---|---|
| Temperature range | -15~100℃ |
| Temperature Fluctuation | ±0.02℃ |
| Temperature Resolution | 0.01℃ |
| Tank Volume | 6L |
| Tank opening Dimensions | 145*145*150mm |
| Temperature control mode | Microcomputer temperature control, self-tuning PID adjustment |
| Circular mode | Internal/external circulation |
| Heating power | 800W |
| Dimensions | 220*410*580mm |
| Flow rate of pump | 10L/min |
| Net Weight of the whole machine | 23Kg |
| Low level protection | Yes. |
FangRui HDC1506L Thermostatic Bath Packing list
host x1, cover x1, silicone tube x1, power cord x1, manual, certificate of conformity, warranty card x1.
[Note] Because the manufacturer's packaging may be updated or upgraded, the detailed packaging list shall be subject to the latest standard configuration of the manufacturer.
FAQ
- How to assemble and use Thermostat
- Thermostat structure and working principle introduction
- Viscosity Determination of Eccentric Falling Ball Determination
- Instructions for assembly and use of Thermostat
- 0.01 ℃ High Precision Low Temperature Thermostat


HDC1506L