Search
    Cart(0)
      总价
      前往结算
      您的购物车是空的 : )
      Home> STIP-scan> LeiCi> LeiCi DZS-706 Multiparameter Water Quality Analyzer

      LEICI DZS-706 multiparameter Water Quality Analyzer Measuring temperature/dissolved oxygen/conductivity/PH value

      LEICI DZS-706 multiparameter Water Quality Analyzer is suitable for the detection of temperature value, dissolved oxygen value, conductivity value, pH value and other water parameter values in aquaculture, agriculture and forestry, and other industries.

      Brand
      SKU: NB001165
      China LeiCi
      Model
      VS
      Genuinegoods
      • LeiCi DZS-706 Multiparameter Water Quality AnalyzerOverview

      • Specifications
      • Comments
      • FAQs
      DZS-706
      Multiparameter Water Quality Analyzer

      Introduction

      Thunder Magnetic DZS-706 multi-parameter Water Quality Analyzer is a desktop multi-function water quality testing instrument, with a large LCD LCD screen, with backlight display function, the operation language is all Chinese design, with continuous, timing, balanced three measurement modes, can achieve different measurement requirements.DZS-706 multi-parameter Water Quality Analyzer can measure temperature value, conductivity value, dissolved oxygen value, PH value and other parameter values at the same time, the measurement is accurate and reliable, the measurement PH value level is 0.01 grade, and the conductivity is 1 grade.

      Introduction to functional modules

      1. Dissolved oxygen measurement module function

      DZS-706 multi-parameter Water Quality AnalyzerIt has the functions of automatic zero oxygen calibration and full calibration, and can realize two calibration methods: air pressure calibration and salinity calibration.

      2. Conductivity measurement module function

      1)DZS-706 multi-parameter Water Quality AnalyzerIt supports the measurement of conductivity, resistivity, TDS, and salinity.

      2) Automatic calibration, automatic range switching, can calibrate the cell constant or TDS conversion coefficient. Among them, DZB-718 supports 2-point calibration.

      3. PH/ion measurement module function

      1) Automatically identify 10 kinds of buffer solutions, allowing self-built standard solution sets.

      2)DZS-706 multi-parameter Water Quality AnalyzerCapable of automatic 1-5 point calibration, manual/automatic temperature compensation is possible.

      3) Provide a variety of commonly used ion modes such as: H+, Ag+, Na+, K+, NH4+, Cl-, F-, etc., users can also customize ion modes.

      4) Provide a variety of ion concentration units, such as ug/L, mg/L, g/L, mol/L, mmol/L, PX, etc., to achieve automatic switching.

      5) It has a direct concentration measurement mode.

      Features and advantages

      1. Manual/automatic temperature compensation function.

      2.DZS-706 multi-parameter Water Quality AnalyzerIt can measure pX/pH, ORP, conductivity, TDS, salinity, resistivity, dissolved oxygen, saturation, and temperature.

      2. Adopt large screen LCD display, with backlight design, Chinese operation interface.

      4. Support zero-oxygen calibration and fullness calibration.

      5. When the conductivity is measured, the frequency is automatically switched to ensure the measurement accuracy.

      6.DZS-706 multi-parameter Water Quality AnalyzerWith power-off protection, it can protect the data security.

      Live shot

      Detail diagram of the main unit of the DZS-706 multi-parameter Water Quality Analyzer

      VS

      LeiCi DZS-706 Multiparameter Water Quality AnalyzerSpecifications

      LIST VALUE
      Instrument level pH: 0.01 grade; conductivity: 1.0 grade
      Measurement parameters Conductivity Detector, dissolved oxygen, pH/pX, temperature
      pH/PX Measurement range (-2.00~20.00)pH/pX
      ORP Measurement range ±1999.9mV
      Ion concentration Measurement range (0~19990) Optional units: pX, mol/L, mmol/L, g/L, mg/L, ug/L
      Conductivity range 0.000μS/cm~199.9mS/cm
      Resistivity range 0.00Ω·cm~20.00MΩ·cm
      TDS range 0.000mg/L~100g/L
      Salinity range 0.00~8.00%
      DO range 0.00~19.99mg/L
      DO% Measurement range 0.0~199.9%
      Response time ≤ 45s (90% Response at 20 ℃)
      Measurement range -5.0~105.0℃
      pH/pX resolution 0.01pH/pX
      ORP resolution 0.1mV
      EC resolution 0.001μS/cm
      Resistivity resolution 0.01Ω·cm
      TDS resolution 0.001mg/L
      Salinity resolution 0.01%
      DOS resolution 0.01mg/L
      DO% resolution 0.10%
      Temperature Resolution 0.1℃
      pH/pX intrinsic error +/- 0.01pH +/- 1 word
      ORP intrinsic error ±0.1%FS
      EC intrinsic error ±1.0%FS
      Resistivity intrinsic error ±1.0%FS
      TDS intrinsic error ±1.0%FS
      Salinity intrinsic error ±0.2%
      DO intrinsic error ±0.30mg/L
      DO% intrinsic error ±10.0%
      Temperature intrinsic error +/- 0.3 ℃ +/- 1 word
      Temperature compensation range Automatic, PH: (0.0~100.0) ℃; Conductivity Detector: (0.0~100.0) ℃; dissolved oxygen: (0.0~40.0) ℃
      Power Supply DC universal Power Supply (9V DC, 800mA)
      Dimensions 280*215*93mm
      Weight 1kg

      LeiCi DZS-706 Multiparameter Water Quality Analyzer Packing list

      Multiparameter analyzer x1, DJS-1C type Conductivity detector electrode (platinum black) x1, E-201-C type pH composite electrode x1, DO-958-S type dissolved Oxygen electrode x1, T-818-B-6 type temperature electrode x1, REX-3 type multifunctional electrode holder x1, pH standard buffer pH4,7,9 each x5, general power supply (9VDC, 800mA, internal positive and external negative) x1, communication disc x1, USB communication cable x1, manual x1, certificate x1

      [Note] Because the manufacturer's packaging may be updated or upgraded, the detailed packaging list shall be subject to the latest standard configuration of the manufacturer.

      FAQ

      [Q]
      DZS-706 measurement TDS value, how to choose Conductivity detector electrode?
      [A]
      Hello, when measuring high TDS, a large constant Conductivity detector electrode is generally used. When TDS ≥ 10.00 g/L, an electrode with an electrode constant of 5 or 10 needs to be used.
      [Q]
      How does DZS-706 Water Quality Analyzer operate? Do you have any instructions?
      [A]
      Hello, each instrument is equipped with an operation manual. You can also download the "operation manual" on this product page to read.
      [Q]
      What parameters can the DZS-706 Water Quality Analyzer test?
      [A]
      Hello, DZS-706 supports the measurement of several parameters such as PH/PX, Conductivity detector, dissolved oxygen, and temperature. It needs to be equipped with corresponding modules to perform measurement operations.
      [Q]
      Is there anything to pay attention to when the DZS-706 first operates the test Conductivity detector value?
      [A]
      Hello, Conductivity detector electrode is used for the first time or after long-term storage, soak the electrode in absoluteethanol for 1 minute, and then fully clean the Conductivity detector electrode with deionized water.
      [Q]
      Do I need to set atmospheric pressure and salinity values during the DZS-706 test?
      [A]
      Hello, no settings are allowed. The default atmospheric pressure value of this Water Quality Analyzer is 101.3KPa, and the default salinity value is 0.0g/L.
      LeiCi
      LEICI DZS-706 multiparameter Water Quality Analyzer Measuring temperature/dissolved oxygen/conductivity/PH value