      Home> Coating Thickness Gauge> TIMES> TIMES F10 probe
      TIMES F10 probe

      TIMES F10 Coating Thickness Gauge: 0~10000um

      Brand
      SKU: NB013331
      China TIMES
      Model
      VS
      TIMESLOGO

      TIMES

      Information
      Genuinegoods
      • TIMES F10 probeOverview

      • Specifications
      • Comments
      • FAQs
      F10
      probe

      Times Aerospace F10 Coating Thickness Gauge Probe is an optional probe for TT260 Coating Thickness Gauge, with a thickness measurement range of 0~10000 microns, a resolution of 10 microns, and a minimum critical thickness of 2mm, which is a magnetic probe.

      Times Aerospace F10 coating Thickness Gauge probe display diagram

      Probe parameter table

      Probe model

      F400

      F1

      F1/90

      F10

      N1

      CN02

      Operating Principle

      magnetic induction

      Eddy current

      Measuring range (um)

      0-400

      0-1250

      0-10000

      0-1250 (0-40um chrome plating on copper)

      10-200

      Low Resolution (um)

      0.1

      0.1

      10

      0.1

      1

      Indication error

      One Point Calibration (um)

      ±[3%H+1]

      ±[3%H+10]

      ±[3%H+1.5]

      ±[3%H+1]

      Two-point calibration (um)

      ±[(1~3)%H+0.7]

      ±[(1~3)%H+1]

      ±[(1~3)%H+10]

      ±[(1~3)%H+1.5]

      -

      Measurement conditions

      Minimum radius of curvature (mm)

      Convex 1

      Convex 1.5

      Straight

      10

      3

      Straight

      Diameter of the minimum area of the substrate (mm)

      ф3

      ф7

      ф7

      ф40

      ф5

      ф7

      Minimum Critical Thickness (mm)

      0.2

      0.5

      0.5

      2

      0.3

      Unlimited

      VS

      TIMES F10 probeSpecifications

      LIST VALUE
      Measurement principle magnetic Induction
      Measurement range 0~10000μm
      margin of error One point calibrated (μm ) : +/- [ 3% H + 10], two point calibrated (μm ): +/- [( 1~ 3) % H + 10]
      resolution 10μm
      Minimum measurement surface Φ40mm
      Minimum radius of curvature 10mm
      thinnest substrate 2mm
      probe F10

      TIMES F10 probe Packing list

      F10 probe X1, manual X1, certificate X1, warranty card X1

      [Note] Because the manufacturer's packaging may be updated or upgraded, the detailed packaging list shall be subject to the latest standard configuration of the manufacturer.

      FAQ

      [Q]
      Is the price listed on the website the final cost for overseas purchases?
      [A]
      No, it isn't. This is the ex-factory price of the product in the Chinese Mainland. In addition to this price, buyers from different countries need to account for additional costs such as shipping fees, tariffs, and others. For specific regional quotes, please consult our customer service or fill out a quotation form.
      [Q]
      What is the specific process for purchasing the product?
      [A]
      First, please click on the quotation form on the website and fill in detailed information. Our customer service personnel will then send you a quotation based on your specific country and region, confirming payment and delivery methods. Following this, we proceed with contract signing, payment, and shipment. You can then await delivery.
      [Q]
      What should I do if the instrument I purchase needs repair?
      [A]
      After-sales repair for overseas instruments is an unavoidable issue. Our repair terms stipulate that the sender is responsible for the shipping and insurance costs incurred for the repair. Repairs within the warranty period are free, while repair costs beyond the warranty period are determined based on the actual situation.
      TIMES
      TIMES F10 Coating Thickness Gauge: 0~10000um
      Service