LABO LB-CW16 Corrugating meter high-precision PID thermostatic control
Adopt high-precision PID thermostatic controller, temperature adjustable range RT +~ 200 ℃, working Rotation speed 4.5 +/- 1rpm, with automatic memory and parameter setting function, suitable for corrugated base paper strength test.
LaboIntroduction
Labo LB-CW16 flute Tester, also known as flute machine, is used to measure the flat and vertical compressive strength of corrugated base paper. During the test, the sample is corrugated on the groove meter, and then glued into a single-sided corrugated or clamped in the vertical compressive strength clamp with adhesive tape, and compressed on the compression Tester to measure its strength.
Meets standards
GB/T1061、ISO7263、TAPPI T809
Features
Labo LB-CW16 groove gauge is controlled by a single-chip microcomputer, using a high-precision PID thermostat, digital display of the actual temperature and set temperature, and an over-temperature safety protection device. It has the functions of automatic memory, parameter setting and adjustment, and adopts precision gear transmission mechanism.
Pictures

Product testing
Cut the sample of length (152.0±0.2) mm, width (12.7±0.1) mm strictly from the sample with sampling equipment, the long side is the longitudinal of the sample, the sample is put into the two sticks of the pre-heated groove meter (heated to 175 °C±8 °C), the sample is fluted, the sample is fluted, the sample is prepared according to the requirements of the test regulations with a flat pressure comb tooth or a vertical pressure clamp, and then placed in the center position of the pressure plate under the compression Tester, and the maximum force that the sample bears when the sample is crushed is read, and the total measurement is carried out 10 valid data, expressed in their arithmetic mean, to the accuracy of 1N.
Labo LB-CW16 Corrugating MeterSpecifications
| LIST | VALUE |
|---|---|
| Rotation speed | 4.5±1rpm |
| Rotation speed error | ±0.2rpm |
| Temperature adjustable range | RT+~200℃ |
| Operating temperature | 176℃ |
| Temperature Resolution | 1℃ |
| An indication of temperature error | ±1% |
| Variability of temperature indication | ≤1% |
| Working pressure adjustable range | 100±10N |
| Set pressure and Working pressure error | ±5% |
| Grooved wheel outer diameter | 228.5±0.5mm |
| Flute wheel thickness | 16±1mm |
| number of teeth | 84. |
| tooth pitch | 8.50±0.05 mm |
| Tooth width | 19mm |
| tooth height | 4.75±0.05mm |
| Gear tooth form | A-type flute |
| Spring tension | 100±5N |
| steel plate | (150×25×0.8)mm |
| Power supply Voltage | AC220V +/- 22V 50Hz 0.5A should be reliably grounded |
| Dimensions | 570*400*330mm |
| Weight | 52kg |
| Rotation speed error | ±1.0RPM |
| tooth height | 4.75±0.05mm |
| number of teeth | Nine. |
Labo LB-CW16 Corrugating Meter Packing list
host X1, manual X1, certificate X1, warranty card X1
[Note] Because the manufacturer's packaging may be updated or upgraded, the detailed packaging list shall be subject to the latest standard configuration of the manufacturer.
FAQ
- Optical inspection laboratory for coral reefs in Bermuda
- Application of Julabo Heating and Refrigeration Circulator in Petrochemical Industry
- Application of Julabo Heating and Refrigeration Circulator in Automotive Industry
- Application of Julabo Heating and Refrigeration Circulator in Semiconductor, Photovoltaic Photovoltaic Industry
- Application of Julabo Oscillating Bath in Pharmaceutical and Medical Industry
- Application of Julabo Heating and Refrigeration Circulator in Food Industry
- German Julabo heating thermostatic bath model comparison table


LB-CW16