      Home> Coating Thickness Gauge> HBO> HBO HC-220 Coating Thickness Gauge

      EDDY current Coating Thickness Gauge HC-220 Hypertherm nonmagnetic Coating Thickness Gauge N type probe, can measure chrome plating on copper

      Measurement principle: eddy Current, Measurement range 0~1250μm, error range +/- 3%, resolution 0.1μm | 1μm, is a nonmagnetic Thickness Gauge, suitable for measurement of the thickness of nonconductive coatings on nonmagnetic metal substrates, can also be used for measurement of chromium plating on copper, measurement readings are displayed through the LCD screen, simple and convenient operation.

      Brand
      SKU: NB001381
      China HBO
      Model
      VS
      Genuinegoods
      • HBO HC-220 Coating Thickness GaugeOverview

      • Specifications
      • Comments
      • FAQs
      HC-220
      Coating Thickness Gauge
      • EDDY current Coating Thickness Gauge HC-220 Hypertherm nonmagnetic Coating Thickness Gauge N type probe, can measure chrome plating on copper
      • EDDY current Coating Thickness Gauge HC-220 Hypertherm nonmagnetic Coating Thickness Gauge N type probe, can measure chrome plating on copper Figure 1
      • EDDY current Coating Thickness Gauge HC-220 Hypertherm nonmagnetic Coating Thickness Gauge N type probe, can measure chrome plating on copper Figure 2
      Measurement principle: eddy Current, Measurement range 0~1250μm, error range +/- 3%, resolution 0.1μm | 1μm, is a nonmagnetic Thickness Gauge, suitable for measurement of the thickness of nonconductive coatings on nonmagnetic metal substrates, can also be used for measurement of chromium plating on copper, measurement readings are displayed through the LCD screen, simple and convenient operation.

      Coating Thickness Gauge selector

      Introduction

      The HC-220 Eddy Current Coating Thickness Gauge is a Hypertherm Instrument Eddy Current Coating Thickness Gauge with N-type transducer, a non-magnetic Thickness Gauge designed to measure the thickness of non-conductive coatings on non-magnetic metal substrates (in the range of 0~1250 microns) and can also be used to measure chrome plating on copper in the range of 0~40 microns.HC-220The eddy current coating Thickness Gauge is a silver-white case with only four buttons on the panel, and the measurement readings are displayed on the LCD display for easy and convenient operation.

      Features

      1. HaibaoinstrumentHC-220The eddy current coating Thickness Gauge uses the eddy current thickness measurement method to measure the thickness of the non-conductive covering layer on a non-magnetic metal matrix;

      2. There are two measurement methods: continuous measurement mode and single measurement mode;

      3. Two working methods: direct mode and group mode;

      4. There are five statistics: average, maximum, minimum, number of tests, and standard deviation;

      5. Two methods are used to calibrate together, and the systematic error of the probe can be corrected by the basic calibration method;

      6. It has the function of storage: it can store 500 measured values;

      7. It has the function of deleting: deleting a single suspicious data that appears in the measurement, and all the data in the storage area can be deleted in order to carry out new measurements;

      8. The limit can be set: the measured value outside the limit can be automatically alarmed; and a histogram can be used to perform Fenix on a batch of measurements;

      9. With printing function: it can print measured values, statistical values, limits, and histograms;

      10. It has the function of communicating with the PC: the measured value and statistical value can be transmitted to the PC for further processing of the data;

      11、HC-220Eddy current coating Thickness GaugesIt has the function of power supply under-Voltage indication;

      12. There is a buzzer prompt during the operation;

      13. It has the function of error prompt, and the error prompt is carried out through the screen display or buzzer;

      14. There are two shutdown modes: manual shutdown mode and automatic shutdown mode.

      Product structure diagram

      HC-220The composition of the eddy current coating Thickness Gauge includes: probe connection, communication interface, display, keyboard, and battery compartment. The details are as follows:

      Hypertherm HC-220 Eddy Current Coating Thickness Gauge Outline and Construction Diagram

      Description of LCD display parameters

      Hypertherm HC-220 Eddy Current Coating Thickness Gauge LCD Parameter Description

      Actual picture of the product

      Hypertherm HC-220 Eddy Current Coating Thickness Gauge Exterior Drawing

      Hypertherm HC-220 Eddy Current Coating Thickness Gauge probe interface diagram

      Hypertherm HC-220 Eddy Current Coating Thickness Gauge Accessories

      Hypertherm HC-220 Eddy Current Coating Thickness Gauge box drawing


      VS

      HBO HC-220 Coating Thickness GaugeSpecifications

      LIST VALUE
      Measurement principle eddy Current
      Measurement range 0~1250μm
      margin of error ±3%
      resolution 0.1μm|1μm
      Minimum measurement surface Φ5mm
      minimum radius of curvature Convex 1.5mm; Concave 3mm
      thinnest matrix 0.3mm
      probe NF split probe
      statistical function Measurement Times, MAX Maximum, MIN Minimum, Average, Standard deviation
      icon function without
      Calibrated Certificate without
      system language Chinese
      Display screen Back view, four-digit reading display, two-line statistical display, Chinese operation
      companion software without
      Weight 200g
      Dimension 140*65*30mm
      Data storage 500 Measured values
      Calibrated mode Manufacturer calibrated, zero calibrated, calibrated foil calibrated
      Operating temperature 0~40℃
      Relative Humidity 20~90%RH

      HBO HC-220 Coating Thickness Gauge Packing list

      host x1, probe x1, standard piece x5, substrate x1, alkaline battery x2, certificate x1, warranty card x1, manual x1, instrument case x1.

      [Note] Because the manufacturer's packaging may be updated or upgraded, the detailed packaging list shall be subject to the latest standard configuration of the manufacturer.

      FAQ

      [Q]
      Ask what kind of thickness gauge is used for the outermost chrome plating of iron copper? Measurement of the thickness of the outermost chrome layer
      [A]
      It is equivalent to testing the chromium plating layer on copper and can be tested by an electrolysis thickness gauge.
      [Q]
      Can HC-220 measure the thickness of galvanized layer on zinc alloy?
      [A]
      Can't measure.
      [Q]
      Are all Coating thickness gauges available for testing in the ISO 2360 standard?
      [A]
      Not necessarily, but this HC-220 Coating thickness gauge can be considered.
      [Q]
      Can the battery be replaced when the HC-220 thickness gauge is turned on? Will it affect it?
      [A]
      Hello, the battery of the thickness gauge cannot be replaced in the power-on state, which will cause damage to the instrument. Please turn off the instrument first for battery replacement.
      [Q]
      Can HC-220 Coating thickness gauge be tested on curved workpieces?
      [A]
      Hello, this thickness gauge is not suitable for curved surface thickness measurement. You can choose a special thickness gauge for curved surface testing.