      Home> STIP-scan> Bante> Bante Bante901P-CN Portable Multi-ParaMeter Water Quality Meter

      Bante901P-CN Portable multiparameter STIP-scan

      CHINA electrode, high performance portable pH Conductivity Meter, contains 4 measurement functions, the instrument is suitable for outdoor measurement of pH, mV, conductivity and total dissolved solids (TDS) of samples.

      Brand
      SKU: NB014741
      China Bante
      VS
      Genuinegoods
      • Bante Bante901P-CN Portable Multi-ParaMeter Water Quality MeterOverview

      • Specifications
      • Comments
      • FAQs
      Bante901P-CN
      Portable Multi-ParaMeter Water Quality Meter
      • Bante901P-CN Portable multiparameter STIP-scan
      • Bante901P-CN Portable multiparameter STIP-scan Figure 1
      • Bante901P-CN Portable multiparameter STIP-scan Figure 2
      CHINA electrode, high performance portable pH Conductivity Meter, contains 4 measurement functions, the instrument is suitable for outdoor measurement of pH, mV, conductivity and total dissolved solids (TDS) of samples.

      Bante901P Portable pH/Conductivity Meter

      Multi-parameter portable pH meter, including 4 measurement functions, the meter is suitable for outdoor measurement of pH/mV, conductivity and total dissolved solids (TDS) value of liquids.

      pH Mode Features

      • Multi-parameter Water Quality Analyzer with large backlit LCD display.

      • 1 to 5 point key calibration, automatic identification of pH standard buffer.

      • Optional pH buffer standard (USA/NIST/DIN) or 2-point custom calibration.

      • Automatic electrode slope display, prompting the user whether to replace the pH electrode.

      Portable pH Meter

      Conductivity/TDS mode functional characteristics

      • 1 to 5 point key calibration, automatic identification of conductivity standard solution.

      • Optional cell constant (K=0.1/1/10), TDS conversion factor, normalized temperature (20 or 25°C), linear and pure water temperature compensation modes.

      common features

      • Automatic temperature compensation to ensure accurate measurement over the full range.

      • Automatic Calibration Reminder to remind users to calibrate the meter on a regular basis.

      • Automatic end point locking, retaining stable measurement values for easy viewing and recording.

      Portable pH Meter

      • Automatic electrode diagnosis, you can check the zero offset, electrode slope and conductivity K value.

      • The system menu can customize 12 functional parameters such as the number of calibration points, stability standard, data lock mode, and automatic shutdown.

      • One key reset (Reset) automatically restores the instrument to the factory default settings.

      • Expanded memory can store up to 500 sets of measurement data.

      • The built-in clock automatically marks the date and time of stored data to comply with GLP regulations.

      • USB communication interface to send stored data to PC.

      • Multi-mode power supply scheme (battery, power adapter, computer's USB port) ensures the smooth application of the instrument.

      VS

      Bante Bante901P-CN Portable Multi-ParaMeter Water Quality MeterSpecifications

      LIST VALUE
      pH Measurement Range -2.000~20.000pH
      pH measurement Accuracy ±0.002pH
      pH resolution 0.1, 0.01, 0.001 pH, optional
      pH calibrating point 1 to 5 points
      pH calibrating solution USA, NIST, DIN, Custom
      pH temperature compensation range 0~100℃
      pH temperature compensation Manual/Automatic
      ORP Measurement Range ±1999.9mV
      mV measurement Accuracy ±0.2mV
      mV resolution 0.1/1 mV, optional
      Conductivity Measurement Range 0~20mS/cm (max 200mS/cm)
      EC measurement Accuracy ±0.5%FS
      EC resolution 0.001, 0.01, 0.1, 1
      Conductivity calibrating point 1 to 5 points
      Conductivity calibrating fluid 10µS/cm, 84µS/cm, 1413µS/cm, 12.88mS/cm, 111.8mS/cm
      EC Temperature Compensation Range 0~100℃, 32~212°F
      EC Temperature Compensation Mode Manual or automatic
      EC Temperature Compensation Coefficient 0.0~10.0%/℃
      EC Temperature Compensation Mode mode Linear or pure water
      Conductivity Detector cell constant K = 0.1, 1, 10, custom
      Standardized temperature 20 ℃, 25 ℃, optional
      TDS range 0~10ppt (max 200ppt)
      TDS measurement Accuracy ±1%FS
      TDS conversion factor 0.1~ 1.0 (default 0.5)
      Temperature Measurement Range 0~105℃, 32~221℉
      Temperature measurement Accuracy ±0.5℃
      Temperature Resolution 0.1℃
      Temperature calibrating point 1 point
      Temperature unit ° C, ° F, optional
      Data stability condition Fast or high Accuracy
      Calibrated reminder 1 to 31 days or closed
      shutdown mode Manual or automatic (10, 20, 30 minutes after no button operation)
      data storage 500 sets
      Communication interface USB
      Connector BNC, 6-pin MINI Connector
      Power Supply 3 AA batteries
      Battery life Approximately 150 hours (when the LCD backlight is turned off)
      Dimensions 170*85*30mm
      instrument Weight 300g
      optional accessories P11 type pH electrode, P11-LiCl type pH electrode, P11-HA type pH electrode, P11-NA type pH electrode
      optional accessories P12 type pH electrode, P13 type pH electrode, P15 type pH electrode, P16 type pH electrode, P18 type pH electrode, P19 type pH electrode, P21 type pH electrode, P22 type pH electrode, 501 type ORP electrode, 502 type ORP electrode
      optional accessories Conductivity Detector electrode of CON-0.1 type, Conductivity Detector electrode of CON-1 type, Conductivity Detector electrode of CON-10 type
      optional accessories TP-10K temperature probe, Conductivity Detector standard solution, pH standard buffer reagent (USA standard), pH standard buffer reagent (NIST standard)
      Packaging Dimensions 340*290*80mm
      Encasement Weight 1.5kg

      Bante Bante901P-CN Portable Multi-ParaMeter Water Quality Meter Packing list

      Host X1, E201-BNC type pH electrode X1, Con-1 type Conductivity detector electrode X1, TP-10K type temperature probe X1, pH standard buffer reagent (pH4.01/7.00/10.01) X1, Conductivity detector standard solution (84uS/cm, 1413uS/cm, 12.88mS/cm) X1, AA type battery X1, plastic carrying case X1, manual X1, certificate X1, warranty card X1

      [Note] Because the manufacturer's packaging may be updated or upgraded, the detailed packaging list shall be subject to the latest standard configuration of the manufacturer.

      FAQ

      [Q]
      How accurate is the measurement pH value of the water quality measurement instrument?
      [A]
      Accuracy of +/- 0.002pH, is a high-precision measurement instrument.
      [Q]
      How to calibrate STIP-scan temperature?
      [A]
      Specific operations: 1. Ensure that the temperature probe is connected to the meter; 2. Press and hold the ℃ key for 3 seconds, the meter enters the temperature calibrate mode, and the screen displays the current temperature value; 3. Press up or down to set the temperature value; 4. Press the enter key to confirm, the calibrate is complete.
      [Q]
      The screen displays --- what is the reason?
      [A]
      There are two possible reasons: 1. The electrode Sensors are too drying, which can be solved by immersing in tap water for 10 minutes; 2. Measured value Excessive range: Check whether the Sensors are clean and not contaminated.
      Bante
      Bante901P-CN Portable multiparameter STIP-scan
      Service