      Home> Conductivity Meter> Bante> Bante Bante950-UK Laboratory conductivity /TDS/salinity/resistivity Meter

      Bante950-UK Laboratory conductivity /TDS/salinity/resistivity meter

      BANTE Bante950-UK laboratory conductivity /TDS/salinity/resistivity meter high-precision laboratory Conductivity Meter, 5-point button calibrate, automatic information prompt, built-in system menu can customize 11 functional parameters, the meter is suitable for measurement of liquid conductivity, total dissolved solids (TDS), salinity and resistivity value, Measurement accuracy: 0.5% F.S.

      VS
      Genuinegoods
      • Bante Bante950-UK Laboratory conductivity /TDS/salinity/resistivity MeterOverview

      • Specifications
      • Comments
      • FAQs
      Bante950-UK
      Laboratory conductivity /TDS/salinity/resistivity Meter

      Bante950 Precision Conductivity/TDS/Salinity/Resistivity Meter

      High-precision laboratory conductivity meter, 5-point key calibration, automatic information prompt, built-in system menu can customize 11 functional parameters, the instrument is suitable for measuring liquid conductivity, total dissolved solids (TDS), salinity and resistivity value, measurement accuracy: 0.5% FS.

      Features

      • High-precision laboratory conductivity meter, equipped with a large backlit LCD display.

      • 1 to 5 point key calibration, automatic identification of conductivity calibration solution.

      • Selectable cell constant (K=0.1/1/10), TDS factor, normalized temperature (20 or 25°C), linear and pure water temperature compensation, seawater practical salinity measurement modes.

      Conductivity TDS salinity meter

      Conductivity TDS salinity meter

      • Automatic temperature compensation to ensure accurate measurement over the full range.

      • Automatic end point locking, retaining stable measurement values for viewing or recording.

      • Automatic electrode diagnosis, you can check the electrode slope value (K).

      • The system menu can customize 11 functional parameters such as the number of calibration points, stability standard, data lock mode, and automatic shutdown.

      • One key reset (Reset) automatically restores the instrument to the factory default settings.

      • Expanded memory can store up to 500 sets of measurement data.

      • The built-in clock automatically marks the time and date of stored data to comply with GLP regulations.

      • USB communication interface to send stored data to PC.

      Conductivity TDS salinity meter

      Conductivity TDS salinity meter

      Applications:

      • Water and Wastewater Treatment

      • Research and Education

      • Environmental monitoring and farming

      • printing and chemical industry

      • Fish farming and hydroponics

      • Beverage industry and other industrial applications

      VS

      Bante Bante950-UK Laboratory conductivity /TDS/salinity/resistivity MeterSpecifications

      LIST VALUE
      Measuring range 0.01~20mS/cm
      resolution 0.001, 0.01, 0.1, 1
      Temperature compensation range 0~100℃
      Power DC5V power adapter
      Dimensions 210*188*60mm
      Weight 1.5kg
      Conductivity Detector measurement Accuracy ±0.5% F.S
      Conductivity Detector calibrated point 1 to 5 point
      Calibrating fluid 10µS/cm, 84µS/cm, 1413µS/cm, 12.88mS/cm, 111.8mS/cm
      Temperature compensation method Manual/Automatic
      temperature compensation coefficient 0.0~10.0%/℃
      Temperature compensation mode Linear or pure water
      Conductivity Detector cell constant K = 0.1, 1, 10, custom
      Standardized temperature 20 ℃, 25 ℃, optional
      TDS range 0~10ppt (Max. 20ppt)
      TDS measurement Accuracy ±1% F.S.
      TDS conversion factor 0.1~ 1.0 (default 0.5)
      Salinity range 0~10ppt
      Salinity measurement Accuracy ±1% F.S
      Salinity measurement mode Practical salinity (psu) or seawater salinity (ppt)
      Resistivity Measurement range 0 ~ 10MZ
      Resistivity measurement Accuracy ±1% F.S
      Resistivity display resolution 0.01, 0.1, 1
      Temperature Measurement range 0~105℃
      Temperature measurement Accuracy ±0.5℃
      Temperature display resolution 0.1℃
      Temperature calibrating point 1 point
      Temperature unit ℃/℉
      data lock Manual or automatic end point locking
      Data stability condition Fast or high precision
      Calibrated reminder 1 to 31 days or closed
      shutdown mode Manual/Automatic
      One-click reset function contain
      data storage 500 groups
      Data Interface USB
      Connector 6-Pin MINI Connector
      Display Customized LCD (130 × 110mm)
      Optional accessories TP-10K temperature probe, Conductivity Detector standard liquid, K10 platinum Conductivity Detector electrode (made in UK)
      Packaging Dimensions 300*240*170mm
      Encasement Weight 1.7kg
      Temperature compensation method Manual/Automatic

      Bante Bante950-UK Laboratory conductivity /TDS/salinity/resistivity Meter Packing list

      Host X1, K10 Platinum Conductivity detector electrode (made in UK) X1, TP-10K temperature probe X1, DC5V power adapter X1, Conductivity detector standard solution (84 µS/cm, 1413 µS/cm, 12.88mS/cm) X1, manual X1, certificate X1, warranty card X1

      [Note] Because the manufacturer's packaging may be updated or upgraded, the detailed packaging list shall be subject to the latest standard configuration of the manufacturer.

      FAQ

      [Q]
      I want to do a test on the ASTMD 5682-95 standard. Is this resistivity meter okay?
      [A]
      Sure
      Bante
      Bante950-UK Laboratory conductivity /TDS/salinity/resistivity meter
      Service
      Focus on online sales of Lab Testing Instruments