Bante903-CN Precision pH/Dissolved Oxygen Meter water quality monitor
pH, Oxidation-reduction potential and dissolved oxygen concentration measurement, -2.000~20.000pH, +/- 1999.9mV, 0.00~20.00mg/L, 0~105 ℃.
Bante-
DDKSIC TEMP-01 Temperature electrode BNC (Q6) conventional aqueous solution
$ 101.00 -
LEICI DGB-423 Optoelectronic Colorimetry Water Quality Analyzer Ammonia Nitrogen
$ 521.00 -
TONGAO TR-131H Water Quality Total Phosphorus Rapid Tester
$ 2123.00 -
LEICI calcium ion electrode calibrated solution, suitable for calcium ion composite electrode
$ 106.00 -
LEICI 780800N09 Hexavalent Chromium Reagent Kit for DGB-404F
$ 158.00
Bante903 Precision pH/Dissolved Oxygen Meter
Multi-parameter laboratory pH meter, including 3 measurement functions, the built-in system menu can customize 11 function parameters, the meter is suitable for measuring the pH, redox potential, dissolved oxygen concentration and saturation of liquids.
pH mode features:
• Multi-parameter Water Quality Analyzer with large backlit LCD display.
• 1 to 5 point key calibration, automatic identification of pH standard buffer.
• Optional pH buffer standard (USA/NIST/DIN) or 2-point custom calibration.
• Automatic electrode slope display, prompting the user whether to replace the pH electrode.
ORP mode features:
• 1-point offset calibration, allowing adjustment of displayed values to known standards.
• Relative and absolute mV modes provide accurate ORP readings.

laboratory pH meter
Functional characteristics of dissolved oxygen mode:
• 1 or 2 point pushbutton calibration using air saturated water or zero oxygen liquid.
• Compensation for salinity and atmospheric pressure to improve the measurement accuracy of the instrument.
General features:
• Automatic temperature compensation to ensure accurate measurement over the full range.
• Automatic Calibration Reminder to remind users to calibrate the meter on a regular basis.
• Automatic end point locking, retaining stable measurement values for easy viewing and recording.
• Automatic electrode diagnosis, you can check the zero offset and electrode slope.
• The system menu can customize 11 functional parameters such as the number of calibration points, stability standard, data lock mode, and automatic shutdown.
• One key reset (Reset) automatically restores the instrument to the factory default settings.
• Expanded memory can store up to 500 sets of measurement data.
• The built-in clock automatically marks the date and time of stored data to comply with GLP regulations.
• USB communication interface to send stored data to PC.
Bante Bante903-CN Precision pH/Dissolved Oxygen MeterSpecifications
| LIST | VALUE |
|---|---|
| pH Measurement range | -2.000~20.000pH |
| pH measurement Accuracy | ±0.002pH |
| pH resolution | 0.1, 0.01, 0.001pH, optional |
| pH calibrating point | 1 to 5 point |
| pH calibrating solution | USA, NIST, DIN, custom |
| pH temperature compensation range | 0~100°C, 32~212°F |
| pH temperature compensation | Manual/Automatic |
| ORP Measurement range | ±1999.9mV |
| mV measurement Accuracy | ±0.2mV |
| mV resolution | 0.1, 1mV, optional |
| mV calibrated point | 1 point (only for R.mV mode) |
| mV calibrated range | ±200mV |
| DO range | 0.00~20.00mg/L |
| DO measurement Accuracy | ±0.2mg/L |
| DO resolution | 0.1, 0.01 mg/L, optional |
| DO temperature compensation range | 0~50℃ |
| DO temperature compensation | automatic |
| Pressure compensation range | 60.0~112.5kPa, 450~850mmHg |
| Salinity compensation range | 0~50g/L |
| DO% Measurement range | 0.0~200.0% |
| DO% measurement Accuracy | ±2.0% |
| DO% resolution | 0.1% |
| Temperature Measurement range | 0~105℃, 32~221℉ |
| Temperature measurement Accuracy | ±0.5℃ |
| temperature resolution | 0.1℃ |
| Temperature calibrating point | 1 point |
| Temperature unit | ℃, ° F, optional |
| Data stability condition | Fast or high precision |
| Calibrated reminder | 1 to 31 days or closed |
| shutdown mode | Manual or automatic (10, 20, 30 minutes after no button operation) |
| data storage | 500 groups |
| Communication interface | USB |
| Connector | BNC, 6-pin MINI Connector |
| Display | Customized LCD (130 * 110mm) |
| Power | DC5V power adapter |
| Dimensions | 210*188*60mm |
| Instrument Weight | 1.5kg |
| Optional accessories | P11 type pH electrode, P11-LiCl type pH electrode, P11-HA type pH electrode, P11-NA type pH electrode, P12 type pH electrode, P13 type pH electrode, P15 type pH electrode, P16 type pH electrode, P18 type pH electrode |
| Optional accessories | TP-10K temperature probe, dissolved Oxygen electrode film cap, dissolved Oxygen electrode filler, pH standard buffer reagent (NIST standard), pH standard buffer reagent (USA standard), P19 type pH electrode, P21 type pH electrode, P22 type pH electrode, DO100 type dissolved Oxygen electrode, |
| Packaging Dimensions | 300*240*170mm |
| Encasement Weight | 1.7kg |
Bante Bante903-CN Precision pH/Dissolved Oxygen Meter Packing list
Host X1, E201-BNC type pH electrode X1, DO100 type dissolved Oxygen electrode (made in the United States) X1, TP-10K type temperature probe X1, pH standard buffer reagent (pH4.01/7.00/10.01) X1, dissolved Oxygen electrode filler X1, dissolved Oxygen electrode film cap X1, USB cable X1, DC5V power adapter X1, manual X1, certificate X1, warranty card X1
[Note] Because the manufacturer's packaging may be updated or upgraded, the detailed packaging list shall be subject to the latest standard configuration of the manufacturer.
FAQ
- ActiveGB/T 693-2024
- ActiveJJF 1714-2018
- ActiveJJF 1565-2016
- ActiveJJF 1441-2013
- ActiveJJG 2104-2024
- ActiveHJ 1147-2020
- Active 096-2007
- ActiveSC/T 7006-2001
- AbolishGB/T 12157-2007
- ActiveGB/T 20245.4-2013
- ActiveGB/T 12157-2022
- ActiveGB/T 1576-2018
- ActiveHJ 2515-2012


Bante903-CN