BANTE TB200 laboratory Turbidity Meter
Multiparameter laboratory Turbidity Meter with Total Suspended Solids Measurement mode, 2 to 7 point calibrated, optional 4 Measurement units, instrument for laboratory applications.
Bante-
INESA WGZ-200B Portable Turbidity Meter 200NTU
$ 444.00 -
Yuefeng SGZ-50B LABOratory Turbidity Meter measuring range 0~50NTU/0.01
$ 260.00 -
JINGQI WGZ-400BJ Portable precision low turbidity instrument
$ 613.00 -
Yuefeng SGZ-2000IT Microcomputer Turbidity Meter 2000NTU/0.001
$ 846.00 -
INESA INESA WGZ-4000 desktop Turbidity Meter 4000NTU
$ 1097.00
TB200 laboratory turbidimeter
Multi-parameter laboratory Turbidimeter with total suspended solids (TSS) measurement mode, 2 to 7 point calibration, 4 measurement units optional, the instrument is suitable for laboratory applications.
Features
• High-performance laboratory turbidimeter with 4.5-inch TFT display.
• For 2 to 7 point calibration, use Formazin standard solution.
• Optional 4 turbidity units, including NTU, FNU, EBC and ASBC.
• TSS conversion factor to ensure accurate measurement of total suspended solids.
• Continuous measurement mode, available for indexing or pairing cuvettes.
• Key data lock, immediately retain the current measurement value.

Turbidimeter
• Password protection to prevent unauthorized calibration and settings.
• Help information tips, like an operation wizard to simplify instrument application.
• Built-in system menu, which can customize 9 functional parameters such as date and time, display resolution, and automatic shutdown.
• One-button reset, automatically restore the instrument to factory default settings.
• Expanded memory can store up to 200 sets of measurement data.
• USB communication interface to send stored data to PC.
Bante TB200 Laboratory Turbidity MeterSpecifications
| LIST | VALUE |
|---|---|
| Measurement method | ISO 7027 nephelometric method (90°) |
| Measuring range | 0 ~ 2000 NTU / FNU, 0 ~ 500EBC, 0 ~ 9999 ASBC |
| resolution | 0.01 (0~100 NTU), 0.1 (100~999 NTU), 1 (999~2000 NTU) |
| Accuracy | ±2% (0~500 NTU), ±3% (501~2000 NTU) |
| Default calibrated point | 0.02, 10.00, 200, 500, 1000, 1500, 2000 NTU |
| Light source/wavelength | Infrared Light Emitting Diodes |
| Receiving element | Silicon photoelectric receiver |
| stray light | < 0.02 NTU |
| Colorimetric bottle | 60 (H) × 25 (D) mm |
| Automatic shutdown | Manual or automatic (after 2 hours of no button operation) |
| Data storage | 200 groups |
| Communication interface | USB |
| Operating temperature | 0~60°C |
| Power | DC12V power adapter |
| Dimensions | 250*177*96mm |
| Instrument net Weight | 1.2kg |
| Optional accessories | Suitable for TB series turbidity meter |
| Packaging Dimensions | 300 (L) × 200 (W) × 190 (H) mm |
| Encasement Weight | 1.7kg |
| TSS Measurement range | Determined by the set TSS conversion factor |
Bante TB200 Laboratory Turbidity Meter Packing list
host X1, sample bottle X1, standard liquid 0.02, 200, 500, 1000 NTUX1, flannel-free X1, DC12V power adapter X1, manual X1, certificate X1, warranty card X1.
[Note] Because the manufacturer's packaging may be updated or upgraded, the detailed packaging list shall be subject to the latest standard configuration of the manufacturer.
FAQ
- ActiveJIS K0801-1986
- ActiveJJF 1700-2018
- ActiveJJF 1571-2016
- Active 100-2007
- ActiveDIN EN ISO 15715-2006
- ActiveGB/T 13531.3-1995
- ActiveGB/T 13200-1991
- ActiveHJ 1075-2019
- ActiveGB/T 18570-2009
- ISO 15715-2003


TB200