HYLOLOGY HY-100 (A) Powder Vibration Density Tester, Powder Density Tester
In line with GB/T 5162-2006/ISO3953:1993 standard, with adjustable amplitude, vibration frequency and frequency, support automatic calculation and printing results, ensure high accuracy test, error ≤ 1%.
hylology-
FRARIV F100 Scott Volumeter, Powder Loose Bulk Density Test
$ 661.00 -
Shangtian Jingyi YM2002J density Electronic Balance Liquid specific gravity Density Meter 200g/0.01g
$ 441.00 -
SDPTOP FA1104J Electronic Density Balance 110g/0.1mg
$ 731.00 -
XIONGFA GP-300E Solid Density Meter 300g
$ 327.00 -
XINGYUN JA203M-L Liquid Density Scale 210g/0.001g
$ 506.00
Introduction
Powder density refers to the mass corresponding to a unit volume of powder. Due to the gaps (or pores) between the particles or inside the particles in the powder, the density of the powder is usually less than the true density of the corresponding substance. The tap density refers to the density of the powder contained in the container after being tapped under specified conditions.
HY-100 powder density Tester is a powder density Tester independently developed and manufactured according to the national standard GB/T 5162-2006/ISO3953:1993 (determination of metal powder tap density) and with reference to the United States Pharmacopoeia, with independent intellectual property rights . The instrument covers all indicators in the national standard GB/T 5162-2006/ISO3953:1993.

Features
HY-100 Powder Density Tester is composed of adjustable speed motor, vibrating components, microcomputer and micro printer. The amplitude and frequency of vibration can be set by yourself. During the vibration process, the vibration component rotates at a small angle to ensure uniform material, and it can automatically calculate and print the test results.
hylology HY-100(A) Powder Vibration Density MeterSpecifications
| LIST | VALUE |
|---|---|
| Sample Weight | ≤ 500.00 g |
| sample Volume | ≤ 250.00 ml |
| Number of single vibrations | 0~99999 times, the number of vibrations can be directly input (when set to 0 times, the result output is "Loose Density") |
| Counting method | photoelectric induction |
| Motor allowable torque | 0.86 N.m |
| result output | built-in printing function |
| Frequency of vibration | 0~300 rpm (continuously adjustable) |
| vibration amplitude | Maximum 15mm, integer 1-15mm adjustable |
| Repeatability Error | ≤1% |
| Accuracy error | ≤1% |
| Input power | 220V (50Hz/60Hz----------------40W |
hylology HY-100(A) Powder Vibration Density Meter Packing list
Host X1, vibration component X1, gasket X1, measuringflask X5, vibration damping pad X5, power cord X1, fuse X5, testtube brush X1, silicone pad X1, warranty card X1, manual X1, certificate X1
[Note] Because the manufacturer's packaging may be updated or upgraded, the detailed packaging list shall be subject to the latest standard configuration of the manufacturer.
FAQ
- ActiveJJF 2094-2021
- AbolishGB/T 5162-2006
- ActiveGB/T 31057.2-2018
- ActiveGB/T 5162-2021
- ActiveGB/T 6609.25-2023
- ActiveGB/T 21354-2008
- AbolishGB/T 5009.2-2003


HY-100(A)