Search
    Cart(0)
      总价
      前往结算
      您的购物车是空的 : )
      Home> Drying time Recorder> Moderner> Moderner GZY-2 Drying Time Recorder
      Moderner GZY-2 Drying Time Recorder

      MODERNER GZY-2 Drying Time Recorder 3/6/12/24/48 hours optional

      Upgraded GZY-1 model with longer Timer Setting, 6 strokes, 304.8mm stroke, track period 3/6/12/24/48h

      Brand
      SKU: NB016950
      China Moderner
      Model
      VS
      Genuinegoods
      • Moderner GZY-2 Drying Time RecorderOverview

      • Specifications
      • Comments
      • FAQs
      GZY-2
      Drying Time Recorder
      • MODERNER GZY-2 Drying Time Recorder 3/6/12/24/48 hours optional
      Upgraded GZY-1 model with longer Timer Setting, 6 strokes, 304.8mm stroke, track period 3/6/12/24/48h

      Instrument Introduction

      MODERNER GZY-2 drying time Recorder is an upgraded version of the GZY-1 model, which has a longer Timer Setting, between 3 and 48 hours. The GZY-2 drying time Recorder uses a stylus to draw across the coating at a constant speed and slowly, and observes the scratch tracks formed on the surface of the coating to evaluate the drying time of the coating. GZY-2 drying time Recorder has six styli, which can test six different products at the same time, saving cost and time, and high efficiency.

      GZY-2 functional structure

      1. GZY-2 is driven by a stepping motor to provide accurate movement control.

      2. There are six scribing needles installed on the carriage, and the movement of the carriage drives the scribing needles to move on the glass sheet coated with wet film.

      3. The GZY-2 drying time Recorder will stop automatically when the journey ends.

      4. Lift up the pressure plate on the lower part of the rear side of the carriage to manually move and reset.

      5. The front end of the GZY-2 drying time Recorder is equipped with a ruler, which can convert the moving distance and moving time.

      6. It has a timer of 3~48 hours and has five different moving speeds to choose from .

      Real shot of GZY-2 drying time Recorder

      Appearance of MODERNER GZY-2 drying time Recorder

      MODERNER GZY-2 drying time Recorder front view

      MODERNER GZY-2 drying time Recorder scale details

      MODERNER GZY-2 drying time Recorder scale details

      operation method

      1. Preparatory work before the test: 

      (1) The test should be carried out under constant temperature and humidity conditions.

      (2) The GZY-2 drying time Recorder should be placed on a flat table away from strong magnetic fields and strong electromagnetic pulse interference.

      2. Operation method:

      (1) Turn on the power and turn on the power switch at the back of the GZY-2 Drying Time Recorder.

      (2) If the carriage is not at the starting position on the left side, lift the pressure plate at the lower rear side of the carriage, and manually move the carriage to reset.

      (3) Install a suitable scribing needle (ball head or pointed head) on the scribing needle clip to keep the scribing needle rod perpendicular to the glass sheet.

      (4) Loosen the set screw on the bracket, remove the glass piece, and prepare a wet film on the glass piece (the wet film should meet the relevant standards or meet the requirements agreed by the buyer and seller).

      (5) Put the glass piece back on the bracket, and properly tighten the set screw at one end.

      (6) Put down the stylus clip so that the stylus is in contact with the new coating layer.

      (7) Select the appropriate speed gear, press the corresponding button, and the carriage starts to move slowly to the right. If you need to change the selection at this time, you need to press the stop button to re-select. After the specified time, the GZY-2 drying time Recorder will automatically stop and give a beeping sound.

      (8) Corresponding to each drying stage of the observed trajectory, evaluate the results. The drying time can be obtained by converting the scale on the edge of the GZY-2 drying time Recorder.

      scale

      Explanation: As shown in the picture, the coating film starts to dry after 2 hours of the test (at the beginning of track ③); it has been hard-dried at 5 hours (at the beginning of track ⑤, the track disappears and the product disappears).

      Precautions for use

      1) After using for a period of time, the screw rod and slide rod in the GZY-2 drying time Recorder should be properly lubricated from the openings on both sides.

      optional parts

      GZY-Film Applicator: wet film thickness 38μ/76μ; or wet film thickness 75μ/150μ; (wet film thickness can also be customized according to customer requirements).

      GZY-coating accessories: suitable for 3×23×340 glass sheets.

      GZY-coating film accessories


      VS

      Moderner GZY-2 Drying Time RecorderSpecifications

      LIST VALUE
      Number of strokes 6-Pin
      itinerary 304.8mm
      trajectory period 3/6/12/24/48h
      input Voltage 220V 50Hz
      Overall Dimensions 490*220*130mm
      Total Weight 8kg
      Glass size 23×340mm
      Scratch clip Can hold ball or pointed scribing needles with outer diameter ≤ φ2

      Moderner GZY-2 Drying Time Recorder Packing list

      Host X1, power cord X1, ball head scribe X6, Glass sheet X6, manual X1, certificate X1

      [Note] Because the manufacturer's packaging may be updated or upgraded, the detailed packaging list shall be subject to the latest standard configuration of the manufacturer.

      FAQ

      [Q]
      This film applicator does not include it, right? Is it actually convenient for us to shave with a wire rod by ourselves?
      [A]
      It is best to use a small square coating, the size is just right, the wire rod will test the technology, it is best to use a small square coating, the size is just right, and the wire rod will test the technology.
      [Q]
      How many tests can be done at the same time?
      [A]
      6 courses
      [Q]
      Does the MODERNER Drying Time Recorder meet the HG/T2003 standard?
      [A]
      The GZY-2 Drying Time Recorder can refer to
      Moderner
      MODERNER GZY-2 Drying Time Recorder 3/6/12/24/48 hours optional
      Service
      Focus on online sales of Lab Testing Instruments