Introduction
TP152 dissolved oxygen monitor is a fully intelligent, multi-functional, high measurement performance of the online dissolved oxygen Tester, the instrument adopts photoelectric technology, low maintenance rate, high reliability, no oxygen consumption, no flow rate requirements, can be measured in hydrostatics, good compatibility, all stainless steel shell shielding can be widely used in the determination of dissolved oxygen in natural water and sewage.

Features
Chinese main menu operation, easy to understand, quick and convenient operation
Professional process and integrated design, digital circuit design is stable and durable
Fluorescence sensors maintain their accuracy in the presence of minor scratches or partial contamination and are easy to clean
The fluorescent cap has a long life and is easy to maintain
No polarization, no correction in general, no stirring
There is no need to change the membrane, no electrolyte to change, no need to polish the anode
Undisturbed by common chemicals such as hydrogen sulfide, carbon dioxide, etc
The standard output signal type is optional, and the alarm relay can be set arbitrarily
Data cycle storage function, automatic removal of overflow data, simple operation, convenient query, power failure data storage time of more than 10 years
TIMEPOWER TP152 Dissolved oxygen analyzerSpecifications
| LIST | VALUE |
|---|---|
| Display | 128 * 64 LCD, Chinese display |
| Measurement range | 0.00~20.00mg/L |
| Accuracy | ±1.5%FS |
| Resolution | 0.01mg/L |
| Repeatability | ≤1% |
| Response time | T90<5min (25℃) |
| Temperature compensation range | 0~60℃ |
| Water sample temperature | 5~60℃ |
| Environmental temperature | 5~45℃ |
| Working Relative Humidity | ≤ 90% RH (non-condensing) |
| Storage temperature | -25~ 55 ℃ (electrode should be higher than 0 ℃) |
| Power Supply | AC (85~ 265) V Frequency (45~ 65) Hz |
| current output | 0~10mA, 0~20mA, 4~ 20mA optional, load < 300Ω |
| Alarm relay | AC220V 3A /DC30V 3A |
| power consumption | ≤15W |
| Protection level | IP65 |
| Power down save | > 10 years |
| Dimensions | 145*120*150mm |
| opening size | 138*138mm |
| Weight | 0.64kg |
TIMEPOWER TP152 Dissolved oxygen analyzer Packing list
Host X1, dissolved oxygen composite electrode X1, main board X1, online shell X1, fine steel wire X1, electrode extension wire X1, electrode film head X1, 1 meter immersion kit X1, 1 meter extension tube kit X1, flange mounting kit X1
[Note] Because the manufacturer's packaging may be updated or upgraded, the detailed packaging list shall be subject to the latest standard configuration of the manufacturer.