| Display mode | 128X64 dot matrix LCD, Chinese display |
| Measurement range | 0.01~4000NTU |
| Accuracy | ±1.5% F.S |
| Resolution | 0.01NTU |
| Repeatability | ±1%F.S |
| Response time | T90<1min (25℃) |
| Temperature compensation range | 0~60℃ |
| Water sample temperature | 5~60℃ |
| Environmental temperature | 5~45℃ |
| Environmental humidity | ≤ 90% RH (non-condensing) |
| Storage temperature | -25~ 55 ℃ (excluding electrode, electrode should be higher than 0 ℃) |
| Power supply Voltage | AC (85~ 265) V Frequency (45-65) Hz |
| Current Output | 0~10mA, 0~20mA, 4~ 20mA optional, load < 300Ω |
| Alarm relay | 2 normally open points can be set arbitrarily, AC220V 3A /DC30V 3A |
| Power | ≤15W |
| Protection level | IP65 |
| Power down save | > 10 years |
| Dimensions | 145*120*150mm |
| opening size | 138 mm×138mm |
| Net Weight of instrument | 0.64kg |
| Measurement range | 10NTU、100NTU、500NTU、1000NTU、5000NTU、10000NTU、20000NTU |
Warranty days: 365 days