      Home> Coating Thickness Gauge> ElektroPhysik> ElektroPhysik MiniTest 70FN Coating Thickness Gauge

      Germany EPK MiniTest 70FN Pocket Coating Thickness Gauge magnetic eddy current dual-use

      Magnetic and eddy current dual-use, built-in probe, automatic identification of substrate, thickness measurement range F: 0~3000μm; N: 0~2500μm, error range +/- 2%, with statistical function, support 1 or 2 point calibrate.

      Brand
      SKU: NB001862
      Germany ElektroPhysik
      VS
      Genuinegoods
      • ElektroPhysik MiniTest 70FN Coating Thickness GaugeOverview

      • Specifications
      • Comments
      • FAQs
      MiniTest 70FN
      Coating Thickness Gauge
      • Germany EPK MiniTest 70FN Pocket Coating Thickness Gauge magnetic eddy current dual-use
      • Germany EPK MiniTest 70FN Pocket Coating Thickness Gauge magnetic eddy current dual-use Figure 1
      • Germany EPK MiniTest 70FN Pocket Coating Thickness Gauge magnetic eddy current dual-use Figure 2
      Magnetic and eddy current dual-use, built-in probe, automatic identification of substrate, thickness measurement range F: 0~3000μm; N: 0~2500μm, error range +/- 2%, with statistical function, support 1 or 2 point calibrate.

      Introduction

      • The MiniTest 70 Series Pocket Coating Thickness Gauge provides fast and accurate measurement of non-magnetic coatings on steel0 ~ 3000 μm

      • Insulating coating on non-ferrous metal substrates0 ~ 2500  μm

      • Automatically recognizes the substrate material

      • Built-in probes

      • Universal measurement methods

      • It has a statistical function

      apply

      Quick and easy non-destructive measurement of coating thickness, available in two models

      MiniTest 70 F: Built-in probe for measuring non-magnetic coatings on iron substrates

      MiniTest 70 FN : Built-in probe, dual-purpose version, measuring non-magnetic coatings on ferrous substrates, non-conductive coatings on non-ferrous substrates

      Features

      The four buttons are clear at a glance;

      Built-in statistical functions: number of readings, maximum value, minimum value, average value, standard deviation;

      No training is required to use.;

      There is a sound prompt for each reading;

      A standard AA battery, when the power is too low, the BAT will be displayed, and the reading error will not be caused by insufficient power

      The dual-purpose built-in probe does not need to select the measurement mode, and the instrument will automatically select the measurement mode according to the substrate material: magnetic or eddy current principle

       Display diagram


      VS

      ElektroPhysik MiniTest 70FN Coating Thickness GaugeSpecifications

      LIST VALUE
      Measurement principle Magnetic Induction + eddy Current
      Measurement range F:0~3000μm;N:0~2500μm
      margin of error ±2%
      resolution 0.5μm
      Minimum measurement surface 30mm
      Minimum radius of curvature Convex 5 mm; Concave 40mm
      thinnest substrate 0.5mm(F);40μm(N)
      probe Built-in probe
      Calibrated mode One point calibrated, two point calibrated
      Unit display μm/mils
      statistical function Measurement times, MAX Maximum, MIN Minimum, Average, Standard deviation
      Power Supply 1 * AA (No. 5) - Battery
      Standards DIN EN ISO 1461, 2064, 2178, 2360, 2808, 3882, ASTM B 244, B 499, D7091, E 376
      Operating temperature -10~60℃
      Weight Approx. 80 g
      Dimension Approx. 157 mm long, φ 27 mm
      signal processing Integrated 32-bit signal processing (SIDSP ®)
      Repeatability ± (1 μm + 1 % )
      Storage temperature -20~70℃

      ElektroPhysik MiniTest 70FN Coating Thickness Gauge Packing list

      MiniTest 70FN host × 1, iron and aluminum plates, calibrated foil, operating instructions, tether, Encasement box

      [Note] Because the manufacturer's packaging may be updated or upgraded, the detailed packaging list shall be subject to the latest standard configuration of the manufacturer.

      FAQ

      [Q]
      EPK 70FN Coating thickness gauge measurement Film thickness accuracy?
      [A]
      Hello, the measurement accuracy is 1.5μm ± 2%.
      [Q]
      Is the MiniTest 70FN an iron and non-iron based thickness gauge?
      [A]
      Hello, yes.
      [Q]
      Do you have the pocket Coating thickness gauge involved in ASTMB499?
      [A]
      The MiniTest 70FN Pocket Coating thickness gauge will do
      ElektroPhysik
      Germany EPK MiniTest 70FN Pocket Coating Thickness Gauge magnetic eddy current dual-use
      Service