JFL QHD pendulum damping Tester The QHD pendulum damping Tester is designed and manufactured according to the professional standard ZBN7 2020-86 and with reference to the international standard ISO1522-1973. Determine the film Hardness of Paints varnish and related products.
JFL-
TQC SP0501 Pendulum Hardness Tester pendulum damping Hardness Tester 110VAC/50/60Hz
$ 10343.00 -
PUSHEN BY-III (KP) Pendulum Hardness Tester Double pendulum Pendulum damping Hardness Tester
$ 1454.00 -
TQC SP0510 Persoz for sp0500/sp0501 Pendulum Hardness Tester
$ 955.00 -
TRUIT TR 5120/P Pendulum Hardness Tester Persoz
$ 1377.00 -
PUSHEN BY-III (S) Pendulum Hardness Tester Double Pendulum Hardness Tester
$ 1392.00
Introduction
QHD pendulum damping Tester Specifications and uses
The QHD pendulum damping Tester is designed and manufactured according to the professional standard ZBN72020-86 and the international standard ISO1522-------1973.
This instrument is used for sampling tests of paints, varnishes and related products to determine the film hardness of paints, varnishes and related products.
Principle
The damping time of the swing attenuation of the swing rod on the paint film is expressed in seconds. The hardness of the paint film is based on the pendulum rod touching the paint film surface at a certain cycle. If the surface is softer, the swing decay will be faster, and vice versa. The slower the principle. It is worth noting that there is no conversion relationship between the test results of the Konig and Persoz pendulums with different structures, masses, sizes, and periods, because the pendulum and the coating The interaction between them also depends on the complex elasticity and viscosity of the coating. Therefore, in the product standard to measure the damping time of a certain paint film, only one pendulum is specified.
Structure: QHD pendulum damping Tester is mainly divided into the following parts.
(1) Instrument base: used to support the test plate and swing rod, there is no vertical support rod on the base, and it is connected with a horizontal platform with a working plane.
(2) Pendulum: Konig or Persoz
(3) A technical device that automatically records the number of required swing angles.
(4) Instrument cover

JFL QHD Pendulum damping TesterSpecifications
| LIST | VALUE |
|---|---|
| Swing bar type | Konig pendulum |
| Konig swing Mass | 200±0.2g |
| Konig swing period | 1.4±0.02s |
| Konig Glass Swing Time | 250±10s |
| Number of swings on Konig Glass | 178±7 |
| Konig start and finish swing angle | 6°/3° |
| Persoz swing Mass | 500±0.1g |
| Persoz swing cycle | 1±0.001S |
| Persoz Glass Swing Time | ≥420s |
| Persoz Glass Swing Count | 420 |
| Persoz start and end swing angle | 12°/4° |
JFL QHD Pendulum damping Tester Packing list
host X1, K pendulum X1, manual X1, certificate of conformity X1, warranty card X1
[Note] Because the manufacturer's packaging may be updated or upgraded, the detailed packaging list shall be subject to the latest standard configuration of the manufacturer.
FAQ
- What are the gloss meters that meet the GB/T9966 standard?
- What are the models of stone gloss meter? What is the difference?
- These gloss meters can be used in paper inspection, understand!
- How to choose the right model for Jinfulun JFL gloss meter?
- Jinfulun JFL-B60 Glossiness Meter instruction manual
- Study on uncertainty analysis of Moisture meter measurement results by drying method
- ActiveISO
- ActiveNF
- ActiveEN ISO 1522-2006
- ActiveNF EN ISO 1522-2007
- ActiveHG/T 3369-2003
- ActiveHG/T 3349-2003
- ActiveJB/T 9386-2017
- ActiveASTM D4366-2021
- ActiveISO 1522-2022
- ActiveGB/T 1730-2007
- AbolishJB/T 9386-1999
- AbolishGB/T 1730-1993
- ActiveGB/T 1730-2007
- AbolishGB/T 1730-1993


QHD