      Pushen Ford-2 #Ford Viscosity Cup Aluminum single cup

      Viscosity range 25~ 120cSt, outflow time 30~100s, flow hole diameter 2.53mm, capacity 100ml, cup height 87mm, cup width 55mm, cup body material cathode aluminum oxide, filter material brass, applicable standard ASTM D1200

      Brand
      SKU: NB019428
      China Pushen
      Model
      VS
      Genuinegoods
      • Pushen Ford-2# Ford Viscosity CupOverview

      • Specifications
      • Comments
      • FAQs
      Ford-2#
      Ford Viscosity Cup
      • Pushen Ford-2 #Ford Viscosity Cup Aluminum single cup
      • Pushen Ford-2 #Ford Viscosity Cup Aluminum single cup Figure 1
      • Pushen Ford-2 #Ford Viscosity Cup Aluminum single cup Figure 2
      Viscosity range 25~ 120cSt, outflow time 30~100s, flow hole diameter 2.53mm, capacity 100ml, cup height 87mm, cup width 55mm, cup body material cathode aluminum oxide, filter material brass, applicable standard ASTM D1200

      The Ford (Ford) cup is made according to ASTM D 1200, D333, D365 of the American Society for Testing and Materials paint and raw material standards. It is a convenient viscometer for measuring the viscosity of inks, coatings, paints, etc. It is crafted from a good aluminum cup with a small hole in the bottom and a capacity of about 100ml.

      The Ford (Ford) cup measures the viscosity of the sample by measuring the time required for a certain volume of the sample in the aluminum cup to flow out of the small hole at the bottom.

      Front View of Pushen/Pushen No. 2 Ford Cup

      Details of Pushen/Pushen No. 2 Ford Cup

      Appearance of Pushen/Pushen No. 2 Ford Cup

      Pushen/Pushen No. 2 Ford Cup interior details

      Equipment and Utensils

      1. Stopwatch: Accuracy 0.1 (self-provided)

      2. Ford cup holder (optional)

      Instructions

      1. Before use, use a suitable solvent to wipe the inner wall of the cup (note: please pay special attention to cleaning the small hole of the cup, twist it into a rope with soft paper, and pull it repeatedly in the outflow hole); then dry it in the air or Blow dry with cold air, and do not allow any residual liquid from the past measurement liquid to adhere to the cup or the outflow port.

      2. Select an appropriate cup number to control the outflow time between 20 and 80 seconds (see Specifications)

      3. Stir the test solution evenly, filter it in a sieve with no less than 567 holes/square centimeter, and adjust the temperature of the sample and the outFlow Cup to (23±0.5°C) or another agreed temperature

      4. Adjust the Ford cup bracket to make the upper horizontal bubble center.

      5. Pour the sample into the Ford cup, and block the outflow hole with a finger (if it is a corrosive liquid, you can use a baffle to block the outflow hole). After filling, scrape the cup flat with a metal or glass plate, and remove the excess sample. Scrape the sample into the groove on the edge of the Viscosity Cup, and put the receiving cup in place.

      6. Let go of your finger or push the baffle, the sample flows out vertically, and start the stopwatch at the same time. Stop timing when the first breakpoint occurs, and the number of seconds of outflow time at this time is the viscosity value of the sample to be tested.

      7. Repeat the measurement three times, and take the average value as the final reading.

      8. Immediately clean it with a suitable solvent. Never use metal cleaning tools or wire. If the orifice is soiled with dried deposits, soften it with a suitable solvent and clean it carefully, eg by pulling a soft cloth through the orifice.

      Four. Precautions for maintenance

      Due to the drop and collision of the Ford Cup, the small hole will be damaged or deformed, resulting in measurement errors, so care must be taken during use.

      calibration

      1. The instrument should be calibrated regularly after each year of use (the diameter of the orifice has been calibrated in the factory with Newtonian oil that can be traced back to NIST oil before leaving the factory, and the user does not need to calibrate it within one year after purchase)

      2. Calibration needs to use standard silicone oil in accordance with the relevant national Viscosity Cup measurement regulations.

      other

      The results obtained for a particular dipping cup are expressed in "seconds" at a specified temperature. To convert "seconds" to centistokes, see ASTM D 4212.

      VS

      Pushen Ford-2# Ford Viscosity CupSpecifications

      LIST VALUE
      Viscosity range 25~120cSt
      outflow time 30~100s
      Flow hole diameter 2.53mm
      Capacity 100ml
      Cup height 87mm
      Cup width 55mm
      Cup body material Aluminum cathode oxidation
      Filter material Brass
      Applicable standard ASTM D1200
      Environment temperature 23±0.2℃
      Certificate without
      Weight 292g

      Pushen Ford-2# Ford Viscosity Cup Packing list

      Viscosity cup X1

      [Note] Because the manufacturer's packaging may be updated or upgraded, the detailed packaging list shall be subject to the latest standard configuration of the manufacturer.

      FAQ

      [Q]
      Can this PUSHEN Ford viscosity cup correspond to ASTMD1200?
      [A]
      Yes, please refer to the NBchao standards page for specific standards.
      [Q]
      Epoxy resin Coating viscosity measurement with what material Viscometer is better?
      [A]
      Measurement Coating viscosity generally can choose viscosity cup or Viscometer, viscosity cup mainly measures relative viscosity, Viscometer measures dynamic viscosity. In general, in addition to thixotropic Coating, Coatings under 150s can be coated with 4 cups, and Coatings under 20 can be coated with 1 cup. In addition, Yehui uses ISO flow cup to measure Newton-type liquid Coating viscosity, Stormer Viscometer to measure architectural coatings viscosity, and for all brush coatings, rotational viscometer or Cone and Plate Viscometer measurement can be used.