| Determination method | Rapid catalytic method (chromium method) |
| Measurement range | 0~150Mg/L low range 100~1000mg/L high range (> 1000mg/L diluted determination); |
| measurement error | 5~ 100 mg/L, absolute error ≤ +/- 5 mg/L; 100 mg/L~ 1000 mg/L, relative error ≤ +/- 5 |
| Digestion environment | 165 +/- 1 ℃; digestion time: 15 minutes; |
| Measurement time | Measure 9, 12 or 25 water samples simultaneously in 30 minutes; (the number of water samples varies according to different configurations) |
| An indication of temperature error | <± 1℃ |
| Uniformity of temperature field | ≤ 2℃ |
| An indication of time error | ≤ ± 2% |
| record storage | 20 measurement results can be stored |
| Environmental temperature | (5~40)℃ |
| Working Relative Humidity | Relative Humidity < 85% (non-condensing) |
| Power Supply | AC220V±10% / 50Hz |
Warranty days: 365 days