TOKI TOKI TVE-25H cone-plate viscometer viscosity up to 518500mP.s
The TSE-25H cone-plate viscometer has three range gears of H, R, and U, and the swirl/spin Rotation speed is 0.01~ 100.0rpm. The TSE-25H cone-plate viscometer has automatic stop time and viscosity functions.
TOKI-
HUINUO LVDV-2H digital display swirl/spin high temperature Viscometer 1 million mPa · s 250 ℃
$ 2154.00 -
Fungilab Smart L Rotational Viscometer 6 million cP
Login -
NIRUN NTV-AI250 touch high temperature Rotational Viscometer molten fluid viscosity measurement
$ 3713.00 -
YUEPING NDJ-9S Digital Viscometer 6000000mPa · s
$ 638.00 -
FANGRUI NDJ-8S Digital Viscometer 1mPa.S~ 2 million mPa.s High Viscosity
$ 448.00
Introduction
The combination of "torque balance servo method" and "pivot-free structure" achieves good linearity and reproducibility of data. Automatic range adjustment derived from dynamic range makes the program measurement function more convenient to use.
The "pivot-free structure" that does not use jewel bearings ensures reliable linearity and reproducibility of viscosity values ââover a wide measurement range.
The dynamic range is the biggest charm of TV-25/35 viscometer.
With one viscometer, the viscosity range can be expanded up to 8 times without having to replace the spindle, and the viscosity can be measured continuously.
Features
âTorque balance servo method
Using torque balance and servo method, zero position measurement method is adopted. Even within a wide measuring range, it is not affected.
âMultiple measurement ranges
The wide dynamic range enables multi-weight range measurement. The measurement range can be expanded up to 8 times, and continuous measurement is possible without replacing the rotor.
âNo pivot structure
A flexible hinge system replaces pivots and gems, resulting in a pivot-free structure.
âRotation speed
Increasing at a speed of 0.1 rpm, any rotation speed can be set between 0.1 and 100.0 rpm.
âAutomatic zero point correction and deviation correction
This automatic feature makes instrument calibration easy.
âMeasurement mode memory
Manual mode, automatic stop (time, viscosity) mode, each mode has 5 memory modes.
âAuto stop
The measurement time and viscosity value can be automatically terminated.
âRemote operation
The viscometer can be remotely operated to start and end measurement using external input signals. In addition, the output signal (collector open circuit) can be used to confirm the status of the viscometer by lighting and beeping.
âUpper and lower limit alarms
When the upper and lower limit settings are exceeded, a buzzer will sound.
âPreheating
The measurement can be started automatically after preheating according to the preset preheating time. The time until the measurement starts is displayed as a countdown.
âData exchange
It is interchangeable with conventional viscometer data and complies with JIS and ISO standards.
plan

TOKI TVE-25H Con-plate viscometerSpecifications
| LIST | VALUE |
|---|---|
| Standard rotor | 1°34′×R24 |
| Rotor speed | 0.01~100rpm |
| step | 1001 steps, step 0.1rpm |
| Measurement mode | Manual, automatic stop time, automatic stop viscosity |
| Measurement memory | 5 ways per mode |
| Viscosity display | %,mPa.s,Pa.s,kPa.s,cP/P |
| Measurement Accuracy | ±1% |
| Repeatability | ±0.2% |
| Measuring temperature | 10~60℃ |
| Data output | RS232C,USB |
| input signal | start signal |
| Output signal | Measurement signal, beep signal, measurement stop signal |
| Environment Operating temperature | 0~40℃ |
| Environment Relative Humidity | Relative Humidity less than 90% (no condensation) |
| Power Supply | AC100- 240V;50-60Hz |
| Power | Less than 20W |
| Rotor material | stainless steel |
| host Volume | 105×220×320mm(W×D×H) |
| Base size | 290*300*430mm |
| Weight | About 9Kg (including bracket) |
| Range file | H、R、U |
| Torque | 718.7μN·m、1437.4μN·m、5749.6μN·m |
| Viscosity range | 6.48~64820mP.s、13.0~129600mP.s、51.9~518500mP.s |
TOKI TVE-25H Con-plate viscometer Packing list
Viscometer X1, certificate of conformity, warranty card, manual
[Note] Because the manufacturer's packaging may be updated or upgraded, the detailed packaging list shall be subject to the latest standard configuration of the manufacturer.
FAQ
- ActiveASTM
- AbolishGB/T 14235.8-1993
- ActiveGB/T 13217.4-2020
- ActiveSH/T 0751-2005
- AbolishGB/T 9751-1988
- ActiveGB/T 9751.1-2008
- ActiveASTM D4287-2000
- ISO 2884-1-1999
- ActiveGB/T 9751.1-2008


TVE-25H