| Operating Mode | Interference filter |
| display | LED LCD |
| Sample absorption and spray Volume | <6mL/min |
| Stability | ≤ 3% within 15s; ≤ 5% within 5min |
| Response time | <8S |
| Detection limit | K:≤0.004mmol/L;Na:≤0.008mmol/L |
| Sexual repeability | ≤ 3% |
| linearity error | K:≤0.005mmol/L ;Na:≤0.03mmol/L |
| test range | K: 0 ~ 100ug / ml; Na: 0 ~ 160ug / ml |
| Filter light transmission characteristics | Potassium: Absolute value of peak wavelength error ≤ 7nm Sodium: Absolute value of peak wavelength error ≤ 5nm |
| Filter light transmission characteristics | Potassium: half width ≤ 15nm Sodium: half width ≤ 15nm |
| Filter light transmission characteristics | Potassium: Background transmittance ≤ 0.5% Sodium: Background transmittance ≤ 0.5% |
Warranty days: 365 days
Non-artificially caused product quality problems, the manufacturer provides 1 year free warranty service.