Qiwei TND-4T Benchtop Coating-4 Viscometer Pure Copper/4mm
Comply with GB/T 1723 standard; brass material with bracket; aperture 4mm; capacity 100ml.
-
CSB Zahn Cup YONGLIDA No. 5 Zahn Cup
$ 128.00 -
TQC VF2181 filter DIN53211 standard 1mm diameter
$ 149.00 -
TQC VF2092 Portable ISO Viscosity Cup aluminum alloy/6mm
$ 420.00 -
TQC VF2025 Viscosity Cup filter, stainless steel 5mm
$ 151.00 -
TQC No. 4 Ford Ford Viscosity Cup ASTM4 Viscosity Cup 4.1mm
$ 482.00
Principle and Calculation
The viscosity measured by the Tu-4 viscometer is the conditional viscosity, which is the time for a certain amount of sample to flow out of a hole with a specified aperture at a certain temperature, expressed in seconds. Convert the obtained sample outflow time (s) into kinematic viscosity value (㎜²/s) with the following formula:
When t<23s, t=0.154μ+11
When 23s≤t≤150s, t=0.223μ+6.0
In the formula:
t——outflow time, s;
μ—kinematic viscosity, mm²/s.
main technical characteristics
1. The Tu-4 viscometer is made of brass, with a capacity of about 100ml. Its geometric dimensions and inner wall roughness meet the requirements of "GB/T 1723 Method for Determination of Viscosity of Paint". The bottom is a stainless steel outflow hole.
2. Correction coefficient K: Multiply the viscosity measured by the calibrated viscometer by the K value, which is the conditional viscosity measured by the viscometer. The correction coefficient K should be within the range of 1±0.03, otherwise the viscometer should be replaced. The viscometer should be calibrated periodically according to the frequency of its use.
3. The difference between the measured value and the calculated value of the outflow time of the standard viscosity liquid with known kinematic viscosity from the Tu-4 viscometer should be within ±3% of the calculated value.
QiWei TND-4T Tu -4 ViscoMeterSpecifications
| LIST | VALUE |
|---|---|
| outflow time | 0~150s |
| Flow hole diameter | 4mm |
| Capacity | 100ml |
| Cup body material | pure copper |
| Applicable standard | GB/T 1723 |
| Correction factor | K=1±0.03 |
QiWei TND-4T Tu -4 ViscoMeter Packing list
Viscosity cup × 1, certificate ×, manual × 1
[Note] Because the manufacturer's packaging may be updated or upgraded, the detailed packaging list shall be subject to the latest standard configuration of the manufacturer.
FAQ
- ActiveJB/T 3078-2015
- ActiveGB/T 1727-2021
- ActiveGB/T 17001.6-2022
- ActiveHG/T 5187-2017
- ActiveHG/T 3369-2003
- ActiveHG/T 3349-2003
- ActiveCY/T 205-2019
- ActiveGB/T 13217.4-2020
- ActiveQB/T 5477-2020
- ActiveJT/T 821.4-2011
- ActiveJT/T 821.3-2011
- ActiveJT/T 821.2-2011
- ActiveJJG 743-2018
- ActiveHG/T 3830-2006
- AbolishGB/T 13217.4-2008
- ActiveQB/T 1046-2012
- ActiveGB/T 1723-1993
- ActiveJC/T 423-1991
- ActiveGB/T 4653-1984
- ActiveHG/T 2661-1995


QiWei
TND-4T
TND-4G