      Home> Coating Thickness Gauge> LANDTEK> LANDTEK CM8826f Coating Thickness Gauge

      LANDTEK CM8826f Coating Thickness Gauge magnetic Induction Coating Thickness Gauge platings/paints

      LANDTEK mid-range practical type, external probe, magnetic Induction thickness measurement; RS232 interface; Single or continuous Measurement mode; Wear-resistant hard metal spring guide sleeve probe; USB and Bluetooth data output can be purchased separately.

      Brand
      SKU: NB000269
      China LANDTEK
      Model
      Stock
      Loading…
      Price
      $ 309.00
      Promotion
      Service
        Service
        The delivery and service By NBchao.Com
        VS
        Genuinegoods
        • LANDTEK CM8826f Coating Thickness GaugeOverview

        • Specifications
        • Comments
        • FAQs
        CM8826f
        Coating Thickness Gauge
        • LANDTEK CM8826f Coating Thickness Gauge magnetic Induction Coating Thickness Gauge platings/paints
        • LANDTEK CM8826f Coating Thickness Gauge magnetic Induction Coating Thickness Gauge platings/paints Figure 1
        • LANDTEK CM8826f Coating Thickness Gauge magnetic Induction Coating Thickness Gauge platings/paints Figure 2
        LANDTEK mid-range practical type, external probe, magnetic Induction thickness measurement; RS232 interface; Single or continuous Measurement mode; Wear-resistant hard metal spring guide sleeve probe; USB and Bluetooth data output can be purchased separately.

        Coating Thickness Gauge selector

        Introduction

        LANDTEK CM8826f coating Thickness Gauge is a magnetic coating Thickness Gauge with an external F probe, which is used to measure the thickness of non-magnetic coatings on magnetic metal substrates, and the measurement is stable and fast.The CM8826f Coating Thickness Gauge is a single-probe Thickness Gauge with both single and continuous measurements, as well as a low Voltage prompt for both manual and automatic shutdowns.The CM8826f coating Thickness Gauge is ergonomically designed to be aesthetically pleasing, aesthetically pleasing, and easy to operate.

        Probe introduction

        The F-probe directly measures the thickness of non-magnetic overlays (e.g., paint, plastic, enamel, copper, aluminum, zinc, chromium, etc.) on the surface of magnetically conductive materials such as steel and nickel. It can be applied to the thickness measurement of electroplating, paint layer, enamel layer, aluminum tile, copper tile, babbitt alloy tile, phosphating layer and paper, and can also be used for thickness measurement of hull paint and attachments of underwater structural parts.

        Features and advantages

        1) Iron-based single-use coating Thickness Gauge.

        2)CM8826f Coating Thickness GaugeThere are two types of measurement options: single and continuous.

        3) Metric/Imperial unit conversion.

        4) CM8826f Coating Thickness GaugeIt has functions such as manual shutdown, automatic shutdown and under-Voltage prompt.

        5) There is a buzzer prompt sound during operation, and the buzzer does not sound when continuous measurement.

        6) With the optional accessory RS232C software and cable, it can communicate with the PC computer to realize the functions of data collection, processing, analysis and printing.

        7) Optional Bluetooth™ data output options are also available.

        8) Split type, provide a variety of probes, more mobile measurement.

        9) The spring-loaded guide sleeve probe with wear-resistant hard metal probe can not only measure on hard or rough surfaces, but also ensure that the probe has a constant tight force and stable sampling value.

        10) The display screen is a 4-digit 10mm LCD screen.

        Applications

        LANDTEK CM8826F coating Thickness Gauge is widely used in manufacturing, metal processing industry, chemical industry, commodity inspection and other testing fields.

        Actual picture of the product

        Introduction to LANDTEK CM8826f Coating Thickness Gauge

        Detail of the display of the LANDTEK CM8826f coating Thickness Gauge

        Detail of LANDTEK CM8826f coating Thickness Gauge probe

        LANDTEK CM8826f coating Thickness Gauge function introduction

        LANDTEK CM8826f Coating Thickness Gauge Hand Grip Part Design

        LANDTEK CM8826f coating Thickness Gauge box drawing

        Selection guide

        CM8826f涂层测厚仪选型参照表

        LANDTEK Coating Thickness Gauge Selection List

        兰泰涂层测厚仪选型列表

        VS

        LANDTEK CM8826f Coating Thickness GaugeSpecifications

        LIST VALUE
        Measurement principle Magnetic Induction
        Measurement range 0~1250μm
        margin of error ±3%
        Resolution 0.1μm|1μm
        Minimum radius of curvature F: convex 1.5mm; concave 25mm
        thinnest substrate 0.3mm
        probe F split probe
        icon function without
        Power Supply 4 * AAA battery
        Calibrated Certificate without
        system language Chinese
        Display screen 4-Position 10mm LCD screen
        supporting software without
        Machine Weight 120G (without battery)
        Machine size 126*65*27mm
        Minimum measurement surface Φ6mm
        Unit display μm/mils
        Voltage prompt Low Voltage prompt
        Operating temperature 0~40℃
        Relative Humidity 10~90%RH

        LANDTEK CM8826f Coating Thickness Gauge Packing list

        host x1, probe (iron-based/eddy current) x1 each, reference block (iron-based/aluminum-based) CTG-BS x1 each, calibrated diaphragm CTG-CFx4, portable box x1, manual x1

        [Note] Because the manufacturer's packaging may be updated or upgraded, the detailed packaging list shall be subject to the latest standard configuration of the manufacturer.

        FAQ

        [Q]
        I want to do a test about the ASTMB499 standard. Is this Coating thickness gauge okay?
        [A]
        Sure
        [Q]
        How to calibrate CM-8826F?
        [A]
        [Zero correction] 1. Press the measurement probe on the reference block (or on the measurement body without coating), and then tap the zero correction key ZERO to zero. When pressing the ZERO key, the measurement probe should not shake on the reference block. At the same time, it should be noted that the probe can only be lifted after pressing the ZERO key, otherwise, the zero correction is incorrect. 2. Lift the measurement probe more than 1cm, and then press the probe on the reference block (or on the measurement body without coating) at a normal speed. Observe the measurement value at the same point on the reference block. If the measurement value is near 0, it means that the zero correction is successful. Otherwise, it should be re-zero. [Calibration fullness] 1. According to the measurement coating thickness, select the appropriate standard diaphragm to calibrate fullness. 2. Now place the standard diaphragm on the reference block (or on the measurement body without coating). 3. Then press the measurement probe on the standard diaphragm, and the measurement value will be displayed on the display. If the measurement value is different from the standard diaphragm, the measurement value can be corrected by adding 1 key or subtracting 1 key. 4. In order to ensure the accuracy of the calibration fullness, it can be verified by multiple measurements of the same standard diaphragm.