| Temperature range | RT+~360℃ |
| Measuring method | visual |
| Temperature Resolution | 0.1℃ |
| Heating rate | 0.2/0. 5/1/1 5/2/3/4/5 ℃/min |
| error of indication | Capillary method: ≤ 200 ℃: +/- 0.4 ℃; > 200 ℃: +/- 0.7 ℃; Hot stage method: room temperature~ 100 ℃: +/- 0.4 ℃; 100 ℃~ 200 ℃: +/- 0.6 ℃; 200 ℃~ 300 ℃: +/- 0.8 ℃ |
| Repeatability | ±0.3℃(1.0℃/min) |
| linear Heating rate error | ±10% |
| Capillary tubing size | Outer diameter is 1.4mm; Inner diameter is 1.0mm; Length is 90mm. |
| Sample height | 3mm |
| Slide size | 76.2*25.4mm |
| Field of view diameter | Æ2.0mm |
| Data storage | 1000 sets |
| Display mode | 7-Inch LCD touch screen |
| Calibration method | Multipoint calibrated |
| Interface | USB (connected to the host PC), U disk |
| Printer model | CK NT-T24S (optional) |
| Power Supply | 220V±22V 50Hz±1Hz |
| Dimension | 310*230*505mm |
| Net Weight of instrument | 4Kg |
| Slide size | 18*18mm |
| Slide thickness | 0.17mm |
| Slide thickness | 0.8mm |
Warranty days: 365 days