      Home> High-Low Temperature Chamber> BenTing> BenTing CBGDW-1000L Low/high Temperature Test Chamber
      BenTing CBGDW-1000L Low/high Temperature Test Chamber

      BENTING CBGDW-1000L low/high Temperature Test Chamber -40 ℃/1000L

      BENTING low/high Temperature Test Chamber CBGDW-1000L adopts fiberglass insulation material, Temperature range -40 ℃~ + 150 ℃, temperature error ≤ +/- 1 ℃, CBGDW-1000L is an experimental high-low temperature Environmental Test Chamber with an content volume of 1000L, heating time 3 ℃/min.

      Brand
      SKU: NB028790
      China BenTing
      Stock
      Loading…
      Price
      Please contact customer service
      Promotion
      Service
        Service
        The delivery and service By NBchao.Com
        VS
        Genuinegoods
        • BenTing CBGDW-1000L Low/high Temperature Test ChamberOverview

        • Specifications
        • Comments
        • FAQs
        CBGDW-1000L
        Low/high Temperature Test Chamber

        Features

        1. Enclosure Material: cold plate baking paint

        2. Inner wall material: good mirror stainless steel plate

        3. Box insulation material: glass fiber

        4. Single hinged door (left hinge, right handle when facing the cabinet), equipped with observation window, lighting, window frame/door frame anti-condensation electric heating device

        5. Control panel: controller display screen, U disk interface, RS232 computer communication interface, power switch, status indicator light and sound and light alarm device

        6. Unit: refrigeration unit, water tray, drain hole, condensing fan, condenser air inlet, water supply tank

        7. Power distribution: IO board, machine tool transformer, ballast, intermediate relay, solid state relay, AC contactor, thermal relay, fuse, under-reverse phase relay, air switch

        8. The control system automatically adjusts the energy-saving Operating Conditions of the refrigerator according to the test conditions

        9. The cooling capacity of the evaporator is switched by the solenoid valve driven by the control system

        BGDW series high-low temperature Test Chamber parameter pair

        modelBGDW-80LBGDW-100LBGDW-150LBGDW-225LBGDW-408LBGDW-800LBGDW-1000L
        -20°C, -40°C, -70°C-20°C, -40°C, -70°C-20°C, -40°C; -70°C-20°C, -40°C; -70°C-20°C, -40°C, -70°C-20°C, -40°C, -70°C-20°C, -40°C, -70°C
        temperature range-20°C: -20°C~+150°C -40°C: -40°C~+150°C -70°C: -70°C~+150°C
        temperature fluctuation≤0.5°C (temperature fluctuation is half of the difference between the highest temperature and the lowest temperature measured at the central point)
        temperature error≤±1°C (the average temperature displayed by the studio temperature controller minus the average temperature measured at the central point)
        temperature uniformity≤2.0°C (The temperature uniformity is the arithmetic mean of the difference between the highest temperature and the lowest temperature measured in each test)
        Heating time3°C/min (non-linear no-load)
        cooling time0.75~1℃/min (non-linear no-load)
        Relative Humidity error±2.5%RH
        Humidifier power2.5KW
        Way of workingThe 0℃~-40℃ model adopts one-stage compression, and the -40℃~-70℃ adopts a two-element cascade refrigeration method
        Refrigeration CompressorsFully enclosed low noise rotary compressor
        condenserAir-cooled type: external fin and internal thread tube heat exchanger (water-cooled type: shell-and-tube heat exchanger)
        Internal volume80L100L150L225L408L800L1000L
        Inner box size W*H*D400*500*400500*500*400500*600*500600*750*500800*850*6001000*1000*8001000*1000*1000
        Outer box size W*H*D900*1580*9501000*1580*9501000*1650*10501100*1800*10501300*1900*11501500*2050*13501500*2050*1550
        Weight KG200230250300400600700
        Do not test productIt is forbidden to test products that may produce flammable, explosive, volatile, highly toxic, corrosive and radioactive substances


        VS

        BenTing CBGDW-1000L Low/high Temperature Test ChamberSpecifications

        LIST VALUE
        Temperature range -40~150℃
        Interior size 1000*1000*1000mm
        Temperature fluctuation ≤0.5℃
        temperature error ≤ +/- 1 ℃ (the average temperature of the displayed value of the studio temperature Controller minus the measured average temperature at the center point)
        Temperature Uniformity ≤2.0℃
        Heating rate 3℃/min
        Cooling rate 0.75~ 1 ℃/min (non-linear no-load)
        Humidity Deviation ±2.5%RH
        Humidifier power 2.5KW
        Operating Mode -40 ℃~ -70 ℃ adopts binary cascade refrigeration method
        Refrigeration system Fully enclosed low noise rotary compressor
        Refrigeration system Air-cooled type: external fin internal thread tube heat exchanger (water-cooled type: shell and tube heat exchanger)
        Interior Volume 1000L
        Dimensions 1500*2050*1550mm
        Weight 700kg
        Prohibit testing products Prohibition of testing of products that may produce flammable, explosive, volatile, highly toxic, corrosive and radioactive substances

        BenTing CBGDW-1000L Low/high Temperature Test Chamber Packing list

        complete-built unit X1, manual X1, certificate of conformity X1, warranty card X1

        [Note] Because the manufacturer's packaging may be updated or upgraded, the detailed packaging list shall be subject to the latest standard configuration of the manufacturer.

        FAQ

        [Q]
        How is the user experience of CBGDW-1000L low/high temperature test chamber?
        [A]
        The CBGDW-1000L is equipped with an observation window and lighting, allowing users to observe the sample status in real time during the testing process, ensuring safe and convenient operation.
        [Q]
        How to ensure the safety performance of CBGDW-1000L?
        [A]
        CBGDW-1000L is equipped with sound and light alarm device and multiple protection measures, including power protection, temperature overrun protection, etc., to ensure safety and Stability during operation.
        [Q]
        How is CBGDW-1000L Applied to Drug Stability Testing in the Pharmaceutical Industry?
        [A]
        CBGDW-1000L can be used for drug Stability Testing in the pharmaceutical industry to simulate the storage and transportation environment of drugs under different temperatures and Humidity conditions, evaluate their physical, chemical and microbial Stability under different environment conditions, and ensure drug mass and safety.
        [Q]
        Meet GB/T10589 low/high temperature test chamber
        [A]
        CBGDW-1000L low/high temperature test chamber, please refer to it.