Search
    Cart(0)
      总价
      前往结算
      您的购物车是空的 : )
      Home> Drying Oven> TaiSiTe> TaiSiTe GP-30BE Drying/fostering dual-use box

      TAISITE GP-30BE drying/foster dual-use box 30L

      TAISITE GP-30BE drying/foster dual-use box adopts double air duct forced convection method, is functional type, adopts liquid crystal dual row PID temperature control, fluctuation foster +/- 0.5 ℃, drying +/- 1 ℃, GP-30BE set drying, baking, wax, heat treatment and bacterial foster in one.

      Brand
      SKU: NB029074
      China TaiSiTe
      Model
      VS
      TaiSiTeLOGO

      TaiSiTe

      Information
      Genuinegoods
      • TaiSiTe GP-30BE Drying/fostering dual-use boxOverview

      • Specifications
      • Comments
      • FAQs
      GP-30BE
      Drying/fostering dual-use box

      Benefits:

      • Drying, baking, wax melting, heat treatment and bacterial culture are all in one box

      • High precision, high efficiency, safety, wide range.

      • Convenient operation, fixed value operation, timed operation, automatic stop. PID thermostat, accurate temperature control.

      • Dedicated function keys enable temperature setting.

      • Auxiliary menu, to achieve over-rise alarm, deviation correction, menu locking.

      • Vertical double air duct air jacket, turbine centrifugal forced air supply, more uniform temperature.

      • Speed-controlled fans enable the use of lightweight, fine, powdery samples.

      • The forced fan heat dissipation and air intake structure makes the maximum working temperature of the fan < 50°C, and the operation is long-lasting for a long time.

      • LCD multi-data full-screen operation display, power failure memory, power failure compensation

      • Overlift preventer

      Detail design advantages

      Figure 1: Heating curve: time unit: minutes, temperature unit: °C

      Figure 2,Multi-functionalAble to control the panel

       Detail Figure 1 of the Tester GP-30BE Drying/Incubation Chamber Details of the Tester GP-30BE Drying/Incubation ChamberFigure 2

                                                                      Figure 1 Figure 2

      Parameter comparison

      ModelheaterContent volumeNumber of partition layers
      GP-30B/BEStainless steel heating tubes30L6 floors
      GP-45B/BEStainless steel heating tubes45L7 floors
      GP-65B/BEStainless steel heating tubes65L9 floors
      GP-125B/BEStainless steel heating tubes125L13 floors
      GP-230B/BENichrome heating wire230L17 floors
      Other parameters
      mannerDual air ducts force convection
      Operating temperature rangeIncubation RT+5~80°C and drying 80~250°C
      Temperature resolution0.1℃
      Temperature fluctuationsIncubation ± 0.5 °C dry ± 1 °C
      Temperature distribution accuracyIncubation± 1.0 °C dry ± 2.5%
      Built-inStainless steel plate
      ExteriorCold-rolled steel plate, surface resistance * quality coating
      Insulating materialsAluminum silicate fiber
      Exhaust outletsInner diameter 28mm*1, top
      Temperature control methodType B: Digital tube double-row PID;BE type: LCD double-row PID
      Temperature setting methodTap the four buttons to set it
      Temperature representationType B: Measured temperature display: 4-digit digital upper display, set temperature display: 4-digit digital lower display
      BE type: Measured temperature display: liquid crystal upper display, set temperature display: liquid crystal lower display
      timer0~9999 minutes (with timed waiting function)
      Run the functionFixed-value operation, timed operation, automatic stop
      Program modeMatching
      Additional featuresDeviation correction, menu button locking, power failure compensation, power failure memory (BE type: Overlift preventer)
      sensorpt100
      Safety devicesOverlift Alarm (BE Type Overlift Arrester)
      clapboard2 pcs
      Partition shelves4 pcs
      Configurations can be added(GPBE type) partition, RS485 interface, printer, Recorder, external communication, remote control, program thermometer, wireless SMS alarm, U disk data storage


      VS

      TaiSiTe GP-30BE Drying/fostering dual-use boxSpecifications

      LIST VALUE
      Temperature range Foster: RT + 5~ 80 ℃; drying: 80~250 ℃
      Temperature Fluctuation A.±0.5℃;B.±1℃
      Temperature Resolution 0.1℃
      Temperature Uniformity Foster +/- 1.0 ℃; drying +/- 2.5%
      Volume 30L
      Interior size 310*310*310mm
      Chamber material stainless steel plate
      Shelf (std.) 2pcs
      Shelves (max) 6pcs
      Load per shelf 15kg
      Timer 0~9999Min (with Timer waiting)
      Power AC220V/3.6A 50/60Hz
      Input Power Foster: 0.3Kw; drying: 0.8Kw
      Dimensions 445*440*690mm
      Net Weight 32kg
      Enclosure Material Cold rolled steel plate, surface resistance painting
      Thermal insulation material aluminum silicate fiber
      Heating mode Stainless steel heating tube
      Exhaust port Inner diameter 28mm * 1, top
      thermostat Liquid crystal dual row PID
      Temperature setting Tap the four-button settings
      Temperature representation Measuring temperature display: liquid crystal upper display; setting temperature display: liquid crystal lower display
      Run function Fixed value operation, Timer operation, automatic stop
      Program Mode optional
      Additional features Deviation correction, menu button lock, power outage compensation, power outage memory, speed regulating fan, overlift preventer
      Temperature sensor pt100
      Packing size 560*560*830mm
      Shelf spacing 35mm
      Gross Weight 37Kg
      Safety Devices Overlift alarm, overlift preventer
      Convection mode Double duct forced convection
      Shelf (std.) 2pcs
      Shelf (std.) 4pcs

      TaiSiTe GP-30BE Drying/fostering dual-use box Packing list

      complete-built unit X1, manual X1, certificate of conformity X1, warranty card X1

      [Note] Because the manufacturer's packaging may be updated or upgraded, the detailed packaging list shall be subject to the latest standard configuration of the manufacturer.

      FAQ

      [Q]
      This box can control the temperature of 185 ° C and the sample capacity is about 30L?
      [A]
      This model is suitable, GP-30BE drying/fostering dual-use box Temperature range 80~ 250 ℃, can hold 30L.
      [Q]
      Can all drying foster dual-use boxes be used for testing in the GB/T 5211.3 standard?
      [A]
      Not necessarily, but this GP-30BE drying foster dual-use box can be considered.
      TaiSiTe
      TAISITE GP-30BE drying/foster dual-use box 30L
      Service