      Home> STIP-scan> JENCO> JENCO 6175 STIP-scan
      JENCO 6175 STIP-scan

      JENCO 6175 Desktop STIP-scan pH/oxidation reduction/ion/temperature test

      0~14/0.01pH, -1999.9~ 1999.9/0 mV; manual/automatic temperature compensation, three-point correction, backlit LCD display.

      Brand
      SKU: NB002946
      United States JENCO
      Model
      VS
      Genuinegoods
      • JENCO 6175 STIP-scanOverview

      • Specifications
      • Comments
      • FAQs
      6175
      STIP-scan
      • JENCO 6175 Desktop STIP-scan pH/oxidation reduction/ion/temperature test
      0~14/0.01pH, -1999.9~ 1999.9/0 mV; manual/automatic temperature compensation, three-point correction, backlit LCD display.

      Introduction

      Ren's 6175 Desktop pH Tester is a new generation of 6173 Star Desktop Tester. Stable and easy to use LCD large display panel, this product provides a resolution of 0.1mV for ISE applications, and the design appearance is more streamlined and simple. This product is RoHS certified.

      Features

      • Enlarge the display panel

      • 0.1mV high resolution

      • Automatic buffer identification (US&NIST buffer equipped)

      • The electrode rate is displayed through the electrode zero point recognition and electrode oblique point recognition functions

      • Built-in buffer temperature coefficient

      • Calibration data can be stored and can be used immediately after booting

      • Temperature-compensated dissolved oxygen data is performed manually or automatically

      • Auto-lock mode with auto-sensing shutdown to keep your data safe

      • Use the power supply as a power mixer or battery

      • This product is RoHS certified

      Product display diagram

      Display diagram of Ren's JENCO6175 desktop water quality Detector in the United States

      VS

      JENCO 6175 STIP-scanSpecifications

      LIST VALUE
      Measurement range 0.00~14.00pH
      Resolution 0.01pH
      Accuracy ±0.01pH
      pH correction solution type USA (4.01, 7.00 & 10.01) or NIST4.00, 6.86 & 9.18), user-selected
      Temperature range of pH correction solution 0~60.0℃
      pH calibrating point 1, 2 or 3 point correction
      Temperature compensation range Automatic/Manual 0.0~100.0 ℃
      End point sensing and locking Yes.
      Impedance >10^13Ω
      Power Supply 6 x 1.5V AAA batteries or 115/230V AC converters
      Environmental temperature 0~50℃
      Protection level IP-54
      Dimensions 230*105*72mm
      Weight Approx. 504g (without battery)
      ORP Measurement range ±1999.9mV
      ORP resolution 1mV
      ORP Accuracy ±0.4mV(±1500.0);±1mV(-1999.9~-1500.0and1500.0~1999.9)
      Measurement range 0.0~100.0℃
      Temperature Resolution 0.1℃
      Measurement Accuracy ±0.2℃
      Shop Accessories Stainless steel temperature probe, Fillable Glass pH composite electrode (conventional measurement), Plastic/Glass ORP composite electrode (Oxidation-reduction potential measurement)
      Shop Accessories Composite electrode (Glass), viscous or greasy objects, strong acid and alkali, composite electrode (Glass) (pure water measurement), plastic shell pH composite electrode (conventional measurement)
      Working Relative Humidity <90%

      JENCO 6175 STIP-scan Packing list

      Host X1, 6005P electrode X1, 007N bracket X1, AC power adapter X1, manual X1, certificate X1, warranty card X1

      [Note] Because the manufacturer's packaging may be updated or upgraded, the detailed packaging list shall be subject to the latest standard configuration of the manufacturer.

      FAQ

      [Q]
      Can JENCO 6175 STIP-scan meet the GB/T26811 standard?
      [A]
      Applicable, you can consult the customer service of the mall for details about the GB/T26811 standard.
      [Q]
      How does JENCO 6175 change the standard solution group?
      [A]
      Hello, to change the standard solution group, please turn off the machine first, then press and hold the "Stand" button and turn it on at the same time. When the key is turned on, another group of correction fluids can be selected.
      [Q]
      What is the reason why STIP-scan JENCO6175 shows "Er3"?
      [A]
      Hello, Er3 is the temperature beyond the range of 0.0~ 60.0 ° C, which can reduce the temperature of the standard solution and make it within this temperature range.
      [Q]
      Does the JENCO6175 STIP-scan replacement electrode need to be recalibrated?
      [A]
      Hello, I need it.
      [Q]
      How to operate Ren's STIP-scan 6175 correction?
      [A]
      Hello, pH6175 can do single-point, two-point or three-point correction (6175-3C is two-point correction). If you want to do two-point or three-point correction, the first point correction needs to be 6.86/7. 00.
      JENCO
      JENCO 6175 Desktop STIP-scan pH/oxidation reduction/ion/temperature test
      Service